Finlay Dewar D, Nugent Chris D, Nelwan Stefan P, Bond Raymond R, Donnelly Mark P, Guldenring Daniel
School of Computing and Mathematics and Computer Science Research Institute, University of Ulster, Belfast, UK.
J Electrocardiol. 2010 Nov-Dec;43(6):606-11. doi: 10.1016/j.jelectrocard.2010.07.004. Epub 2010 Sep 15.
In this study, we assess the effects of electrode placement error on the EASI-derived 12-lead electrocardiogram (ECG). The study data set consisted of 744 body surface potential map (BSPM) recordings. The BSPMs, each of which was made up of 117 leads, were recorded from a mixture of healthy, myocardial infarction, and left ventricular hypertrophy subjects. The BSPMs were interpolated to increase the number of data points in the region of the EASI recording electrodes I, E, and A and the precordial leads. This facilitated 3 experiments. Firstly, recording sites I, E, and A were simultaneously moved ±5 cm vertically, in 0.5 cm increments, from their correct locations. Secondly, recording sites I and A were moved horizontally, again up to ±5 cm, in 0.5 cm increments. Finally, all 6 precordial leads were moved vertically in 0.5 cm increments up to ±5 cm. At each movement step, the resulting 12-lead ECG was compared with the original 12-lead ECG. Root mean square error was determined along with the absolute difference in J-point amplitude. Although the EASI leads were found to be less sensitive to electrode misplacement than the standard precordial leads, it was found that when precordial leads were moved up to ±3 cm vertically, the resulting 12-lead ECG more accurately resembled the original 12-lead ECG than a 12-lead ECG reconstructed from accurately positioned EASI leads. Further work is required to establish the effects of electrode misplacement beyond the ±5 cm limits assessed in this study.
在本研究中,我们评估了电极放置误差对基于EASI导联得出的12导联心电图(ECG)的影响。研究数据集由744份体表电位图(BSPM)记录组成。这些BSPM每份由117导联构成,记录自健康受试者、心肌梗死受试者和左心室肥厚受试者的混合样本。对BSPM进行插值处理,以增加EASI记录电极I、E和A区域以及胸前导联中的数据点数。这便于进行3项实验。首先,记录位点I、E和A从其正确位置开始同时垂直移动±5 cm,每次移动0.5 cm。其次,记录位点I和A水平移动,同样最多±5 cm,每次移动0.5 cm。最后,所有6个胸前导联垂直移动,每次移动0.5 cm,最多移动±5 cm。在每个移动步骤中,将所得的12导联心电图与原始的12导联心电图进行比较。确定均方根误差以及J点振幅的绝对差值。尽管发现EASI导联对电极错位的敏感性低于标准胸前导联,但发现当胸前导联垂直移动至±3 cm时,所得的12导联心电图比由精确定位的EASI导联重建的12导联心电图更准确地类似于原始的12导联心电图。需要进一步开展工作,以确定超出本研究评估的±5 cm范围的电极错位影响。