Suppr超能文献

A brain phantom for studying contrast recovery in emission computerized tomography.

作者信息

Pupi A, De Cristofaro M T, Formiconi A R, Passeri A, Speranzi A, Giraudo E, Meldolesi U

机构信息

Dipartimento di Fisiopatologia Clinica, Università di Firenze, Italy.

出版信息

Eur J Nucl Med. 1990;17(1-2):15-20. doi: 10.1007/BF00819398.

Abstract

A brain phantom is described that is characterized by a high anatomical definition and by the possibility of varying the internal contrast with the use of a single radioactive solution. The experimental work was done with a single-photon emission computerized tomographic (SPET) rotating camera. The phantom was used to study the contrast recovery of both the filtered back-projection and an iterative reconstruction algorithm. Moreover, it was also used to find a cross-calibration factor between activity concentrations in the SPET slices and an external reference.