Calcutt S B, Pritchard T M, Hepplewhite C L, Taylor F W, Werrett S T, Arijs E, Nevejans D
Appl Opt. 1993 Nov 20;32(33):6764-76. doi: 10.1364/AO.32.006764.
We describe the design, the development, and the calibration of a radiometer to measure water vapor abundance in the middle atmosphere, using the solar occultation technique from an orbiting platform. The use of gas cells containing water vapor and carbon dioxide leads to a relatively simple, but sensitive, design. This radiometer is shown to be capable of detecting water vapor in the mesosphere, where, because of dissociation by the Sun, the amounts are very small, of the order of 1 part per 10(6) or less. A flight version of the instrument will be launched in 1992 on board the European Retrievable Carrier space platform, and the data will be employed for a better understanding of the Earth's water budget.
我们描述了一种利用轨道平台上的太阳掩星技术来测量中层大气中水汽丰度的辐射计的设计、研制和校准。使用含有水汽和二氧化碳的气室使得该辐射计的设计相对简单但却很灵敏。结果表明,这种辐射计能够探测到中层大气中的水汽,在中层大气中,由于太阳的光致离解作用,水汽含量非常少,约为百万分之一或更低。该仪器的飞行版本将于1992年搭载欧洲可回收载体空间平台发射升空,所获数据将用于更好地了解地球的水收支情况。