Zacharia T Thomas, Kanekar Sangam G, Nguyen Dan T, Moser Kevin
Department of Radiology, Neuroradiology, Milton S. Hershey Medical Center, Pennsylvania State University, Hershey, PA 17033, USA.
Emerg Radiol. 2011 Apr;18(2):103-7. doi: 10.1007/s10140-010-0908-5. Epub 2010 Sep 21.
The purpose of this study is to retrospectively analyze the effect of z-axis modulation for CT head protocols on patient dose and image quality in patients with acute head trauma and stroke. The study was approved by the Institutional Review Board. We retrospectively evaluated the effect of dose modulation on unenhanced CT head examinations in patients with acute head trauma and stroke. Two series of 100 consecutive studies were reviewed: 100 studies performed without dose modulation, 100 studies performed with z-axis dose modulation. Multidetector 16-section CT was performed sequentially and axial 5-mm-thick slices were obtained from base of skull to vertex. With z-axis dose modulation, the same tube current range was maintained, but a computer algorithm altered the tube current applied to each CT section. For each examination, the weighted volume CT dose index (CTDI (vol)) and dose-length product (DLP) were recorded and noise was measured. Each study was also reviewed for image quality by two independent, blinded readers. The variables (CTDI (vol) and DLP, image quality, and noise) in the two groups were compared by using student t test and Wilcoxon rank-sum test. For unenhanced CT head examinations, the CTDI (vol) and DLP, respectively, were reduced by 35.8% and 35.2%, respectively, by using z-axis dose modulation. Image quality and noise were unaffected by the use of this dose modulation technique (P < 0.004). Utilization of z-axis modulation technique for CT head examination in patients with acute head trauma and stroke offers significant radiation dose reduction while image quality is optimally maintained.
本研究的目的是回顾性分析CT头部扫描方案中的z轴调制对急性颅脑外伤和中风患者的辐射剂量及图像质量的影响。本研究经机构审查委员会批准。我们回顾性评估了剂量调制对急性颅脑外伤和中风患者非增强CT头部检查的影响。回顾了连续的两个100例研究系列:100例未进行剂量调制的研究,100例进行z轴剂量调制的研究。使用16排多层螺旋CT依次进行扫描,从颅底至头顶获取5毫米厚的轴向层面图像。采用z轴剂量调制时,管电流范围保持不变,但计算机算法会改变施加于每个CT层面的管电流。对于每次检查,记录加权容积CT剂量指数(CTDI(vol))和剂量长度乘积(DLP)并测量噪声。两名独立的、不知情的阅片者对每项研究的图像质量进行评估。采用学生t检验和Wilcoxon秩和检验对两组中的变量(CTDI(vol)、DLP、图像质量和噪声)进行比较。对于非增强CT头部检查,采用z轴剂量调制时,CTDI(vol)和DLP分别降低了35.8%和35.2%。使用这种剂量调制技术未影响图像质量和噪声(P<0.004)。对急性颅脑外伤和中风患者进行CT头部检查时采用z轴调制技术可显著降低辐射剂量,同时能最佳地保持图像质量。