Department of Orthopaedics, University College London Hospital, London, UK.
Clin Orthop Relat Res. 2011 Jun;469(6):1677-82. doi: 10.1007/s11999-010-1588-x. Epub 2010 Sep 29.
Pelvic radiographs are helpful in assessing limb-length discrepancy (LLD) before and after THA but are subject to variation. Different methods are used to determine LLDs. As a pelvic reference, both ischial tuberosities and the teardrops are used, and as a femoral reference, the lesser trochanter and center of the femoral head are used.
QUESTIONS/PURPOSES: We validated the different methods for preoperative radiographic measurement of LLDs and evaluated their reliability.
LLDs were measured on full-leg radiographs for 52 patients (29 men, 23 women) with osteoarthritis (OA) of the hip and compared with different methods for measuring LLDs on AP radiographs of the pelvis.
The true LLD varied from -8.0 to 9.1 mm. When the biischial line was used as a pelvic reference, the LLD measured on AP pelvis radiographs was different from the true LLD. No difference was found when the interteardrop line was used as a pelvic reference. There was substantial interobserver agreement when the lesser trochanter was used as a femoral reference (kappa = 0.66-0.70) and excellent interobserver and intraobserver agreement for all other measurements (kappa = 0.84-0.93).
Our data show use of the biischial line as a pelvic reference should be discouraged and the interteardrop line is a better alternative. The center of the femoral head is a more reliable femoral landmark compared with the lesser trochanter.
骨盆射线照相术有助于在 THA 前后评估肢体长度差异(LLD),但存在差异。有多种方法用于确定 LLD。坐骨结节和泪滴都被用作骨盆参考,小转子和股骨头中心被用作股骨参考。
问题/目的:我们验证了术前放射测量 LLD 的不同方法,并评估了它们的可靠性。
对 52 例髋关节炎(OA)患者的全长射线照片进行 LLD 测量,并与骨盆 AP 射线照片上测量 LLD 的不同方法进行比较。
真实的 LLD 从-8.0 到 9.1 毫米不等。当使用双坐骨线作为骨盆参考时,AP 骨盆射线照片上测量的 LLD 与真实 LLD 不同。当使用泪滴线作为骨盆参考时,没有差异。当使用小转子作为股骨参考时,观察者之间的一致性很高(kappa = 0.66-0.70),所有其他测量的观察者之间和观察者内部的一致性都非常好(kappa = 0.84-0.93)。
我们的数据表明,不应鼓励使用双坐骨线作为骨盆参考,而应使用泪滴线。与小转子相比,股骨头中心是更可靠的股骨标志。