[关于在自适应放射治疗治疗计划中的可用性,对有限投影数据(半扫描、半扇形)的锥束CT采集的图像质量和剂量计算准确性进行分析]

[Analysis of image quality and dose calculation accuracy in cone beam CT acquisitions with limited projection data (half scan, half fan) with regard to usability for adaptive radiation therapy treatment planning].

作者信息

Kowatsch Matthias, Winkler Peter, Zurl Brigitte, Konrad Thomas, Schröttner Jörg

机构信息

Institut für Health Care Engineering, Technische Universität Graz.

出版信息

Z Med Phys. 2011;21(1):11-8. doi: 10.1016/j.zemedi.2010.08.002.

Abstract

UNLABELLED

Subject of this study is the question of whether cone beam CT (CBCT) images with reduced projection data are suitable for use in adaptive radiation therapy (ART) treatment planning. For this purpose image quality and dose calculation accuracy depending on imaging modality were analysed. In this context, two CBCT-methods will be indicated having reduced projection data sets: Scans acquired with 200° rotation angle in order to accelerate the CBCT process (half scan), or scans with an asymmetric cone beam and detector offset, used to enlarge the field-of-view (half fan).

METHODS

For three different CBCT-modes (On-Board-Imaging, Varian Medical Systems), two of them based on reduced projection data, and a conventional multidetector CT system, the main image quality parameters were studied. Treatment plans for two phantoms were transferred to all datasets and re-computed to analyse dose calculation accuracy. Furthermore imaging dose was measured for all modalities.

RESULTS

All three CBCT-modes showed similar results with regard to image quality. It was found, that a reduction in projection data does not necessarily involve deterioration in image quality parameters. For dose calculation based on CBCT images, a good agreement with the reference plan was found, with a maximum deviation for the mean dose in regions of interest of 1.1%. Imaging dose was found to be 2.5 cGy and 2.9 cGy for the large-FOV mode and the partial rotation mode, respectively, and 5.4 cGy for the 360°-full fan mode.

摘要

未标注

本研究的主题是具有减少投影数据的锥束CT(CBCT)图像是否适用于自适应放射治疗(ART)治疗计划。为此,分析了取决于成像方式的图像质量和剂量计算准确性。在这种情况下,将指出两种具有减少投影数据集的CBCT方法:以200°旋转角度采集的扫描以加速CBCT过程(半扫描),或具有不对称锥束和探测器偏移的扫描,用于扩大视野(半扇区)。

方法

对于三种不同的CBCT模式(机载成像,瓦里安医疗系统公司),其中两种基于减少的投影数据,以及一个传统的多探测器CT系统,研究了主要图像质量参数。将两个体模的治疗计划转移到所有数据集中并重新计算,以分析剂量计算准确性。此外,测量了所有模式的成像剂量。

结果

所有三种CBCT模式在图像质量方面显示出相似的结果。发现投影数据的减少不一定会导致图像质量参数的恶化。对于基于CBCT图像的剂量计算,与参考计划有很好的一致性,感兴趣区域内平均剂量的最大偏差为1.1%。发现大视野模式和部分旋转模式的成像剂量分别为2.5 cGy和2.9 cGy,360°全扇区模式的成像剂量为5.4 cGy。