Fung Y S, Luk S F
Department of Chemistry, University of Hong Kong.
Analyst. 1990 Sep;115(9):1219-21. doi: 10.1039/an9901501219.
A simple differential-pulse polarographic method using a laboratory-built hanging mercury drop electrode as the working electrode was developed for the determination of sorbic acid in fruit juices and soft drinks. Sorbic acid was extracted from the samples with diethyl ether. After reduction of the ethereal solution to a small volume by direct evaporation, the residual ether was dissolved in the supporting electrolyte (25 ml of acetonitrile + 1 ml of 0.06 M acetic acid + 0.8 g of tetraethylammonium bromide). Peak current was measured at -1.7 V. The working range of the method, without dilution or pre-concentration of the samples, was from 4 to 229 p.p.m. for the original juice and drink samples. The validity of the method was confirmed by parallel determinations using the method of the Association of Official Analytical Chemists and by recovery tests on a large variety of juice samples. Satisfactory recoveries and agreement in results from the two methods were obtained. The recovery and precision (relative standard deviation) of the method were 97 +/- 4 and 100 +/- 3%, respectively, for blackcurrant juice for five determinations.
开发了一种简单的差分脉冲极谱法,使用实验室自制的悬汞滴电极作为工作电极,用于测定果汁和软饮料中的山梨酸。用乙醚从样品中萃取山梨酸。通过直接蒸发将乙醚溶液浓缩至小体积后,将残留的乙醚溶解在支持电解质中(25 ml乙腈 + 1 ml 0.06 M乙酸 + 0.8 g溴化四乙铵)。在 -1.7 V下测量峰值电流。该方法的工作范围(无需对样品进行稀释或预浓缩),对于原汁和饮料样品为4至229 ppm。通过使用官方分析化学家协会的方法进行平行测定以及对大量果汁样品进行回收率测试,证实了该方法的有效性。两种方法获得了令人满意的回收率和结果一致性。对于黑加仑汁进行五次测定,该方法的回收率和精密度(相对标准偏差)分别为97 +/- 4%和100 +/- 3%。