Saysin Suwanna, Liawruangrath Boonsom, Liawruangrath Saisunee
Department of Pharmaceutical Science, Faculty of Pharmacy, Chiang Mai University, Chiang Mai 50200, Thailand.
J Cosmet Sci. 2010 Sep-Oct;61(5):367-76.
A high-performance thin-layer chromatographic method was developed for the determination of ketoconazole. The sample was separated on a silica gel 60 F254 plate and developed in ethanol-acetone-1.0 mol l-1 H2SO4 by means of an automatic multiple-development system. The area of the spot was quantified by a TLC scanner at a wavelength of 298 nm. A linear calibration curve was established over the range of 3-20 μg/ml of ketoconazole, with a correlation coefficient of 0.9992. The relative standard deviations for intraday and interday precisions, for three replicate determinations, were found to be 1.72% and 0.69% for 5 μg/ml and 2.18% and 0.94% for 10 μg/ml of ketoconazole, respectively. The average percentage recoveries of ketoconazole shampoos (Nora, Kenalyn, and Nizoral) and ketoconazole creams (Nizoral, Fungasin, and Ketazon) were found to be 96.10, 97.06, and 99.58, and 96.77, 97.26, and 95.74, respectively. This method has been applied to the determination of ketoconazole in various pharmaceutical dosage forms. Common excipients in formulations do not interfere. This method is simple, precise, accurate, and inexpensive. It should be used for routine analysis.
建立了一种高效薄层色谱法测定酮康唑。样品在硅胶60 F254板上分离,通过自动多次展开系统在乙醇 - 丙酮 - 1.0 mol l-1硫酸中展开。斑点面积通过TLC扫描仪在298 nm波长下进行定量。在3 - 20 μg/ml的酮康唑范围内建立了线性校准曲线,相关系数为0.9992。对于5 μg/ml的酮康唑,三次重复测定的日内和日间精密度的相对标准偏差分别为1.72%和0.69%;对于10 μg/ml的酮康唑,分别为2.18%和0.94%。酮康唑洗发水(诺拉、凯纳林和采乐)和酮康唑乳膏(采乐、夫西地酸和酮唑)的平均回收率分别为96.10%、97.06%和99.58%,以及96.77%、97.26%和95.74%。该方法已应用于各种药物剂型中酮康唑的测定。制剂中的常见辅料不产生干扰。该方法简单、精密、准确且成本低廉。应用于常规分析。