Phys Med Biol. 2010 Nov 7;55(21):6431-43. doi: 10.1088/0031-9155/55/21/006. Epub 2010 Oct 15.
In theory, the application of the two-compartment exchange model (2CXM) to data from a dynamic contrast-enhanced (DCE) MRI exam allows the estimation of the plasma flow, plasma volume, extraction flow and extravascular-extracellular volume. The aim of this paper was to explore whether simulations based on the 2CXM could provide useful information on the trustworthiness of the results. The deviations from the input values of the haemodynamic quantities were estimated for a 'reference tissue' with a clear bi-phasic response and four 'limit tissues' with more challenging 2CXM fitting properties. The impact of the instrumental factors sampling step (T(s)), acquisition window (T(acq)) and contrast-to-noise ratio (CNR) was investigated. Each factor was varied separately, while keeping the other ones at a value above concern. Measurement guidelines to ensure that all deviations fell within a predefined range (±20%) could not be derived, but simulations for fixed T(s) and T(acq) were found to provide a practical tool for studying the error behaviour to be expected from a given experimental set-up and for comparing measurement protocols. At the level of an individual DCE exam, a bootstrap version of the simulation approach was shown to lead to a useful estimate of the errors on the fitted parameters.
从理论上讲,将双室交换模型(2CXM)应用于动态对比增强(DCE)MRI 检查的数据,可以估计血浆流量、血浆体积、提取流量和血管外细胞外体积。本文旨在探讨基于 2CXM 的模拟是否能为结果的可靠性提供有用信息。对于具有明确双相反应的“参考组织”和具有更具挑战性的 2CXM 拟合特性的四种“极限组织”,估计了血液动力学量的输入值的偏差。研究了仪器因素采样步长(T(s))、采集窗口(T(acq))和对比噪声比(CNR)的影响。在保持其他因素值高于关注值的情况下,分别改变每个因素。无法得出确保所有偏差都落在预定范围内(±20%)的测量指南,但发现固定 T(s)和 T(acq)的模拟为研究给定实验设置预期的误差行为以及比较测量方案提供了一种实用工具。在单个 DCE 检查的水平上,模拟方法的引导版本被证明可以对拟合参数的误差进行有用的估计。