Division of Orthopaedic Surgery, McGill University Health Centre, Montreal, QC, Canada.
Clin Orthop Relat Res. 2011 Jun;469(6):1642-50. doi: 10.1007/s11999-010-1630-z.
Metal-on-metal articulations can release substantial amounts of particles containing cobalt and chromium into the surrounding milieu, causing concern for cellular toxicity and adverse local soft tissue reactions. The diameter of the femoral head has been one of the variables that inversely affects wear of metal-on-metal total hip arthroplasty (THA). The oxidative stress of increased metal ions can be measured with serum markers. It is still controversial if larger femoral head diameters decrease wear rates in patients with metal-on-metal THA and if the increased metal ions alter the body's antioxidant status.
QUESTIONS/PURPOSES: We therefore (1) determined whole blood metal ions in patients with small (28 mm and 36 mm) and large (40 mm and 44 mm) diameter femoral heads; (2) measured oxidative stress markers (total antioxidants, nitrotyrosine, and peroxides); and (3) determined whether acetabular version or inclination influenced ion levels.
One hundred four patients were retrospectively studied. We recorded Harris hip scores and UCLA activity scores. All patients were followed at 1 year.
The activity scores were similar in the two groups. There was no difference in metal ion levels or oxidative stress markers between patients with small- or large-diameter femoral heads. Acetabular inclination and anteversion had no effect on the metal ion levels.
The data suggest there is no difference in ion values in patients with large or small metal-on-metal THA and the increased metal ions do not alter the oxidant status of the patient.
金属对金属关节会释放出大量含有钴和铬的颗粒到周围环境中,引起人们对细胞毒性和局部软组织不良反应的担忧。股骨头直径是影响金属对金属全髋关节置换术(THA)磨损的一个变量。增加的金属离子的氧化应激可以用血清标志物来衡量。目前仍存在争议的是,在金属对金属 THA 患者中,较大的股骨头直径是否会降低磨损率,以及增加的金属离子是否会改变机体的抗氧化状态。
问题/目的:因此,我们(1)测定了小(28 毫米和 36 毫米)和大(40 毫米和 44 毫米)直径股骨头患者的全血金属离子;(2)测量了氧化应激标志物(总抗氧化剂、硝基酪氨酸和过氧化物);(3)确定髋臼的倾斜度或倾斜度是否影响离子水平。
对 104 例患者进行回顾性研究。我们记录了 Harris 髋关节评分和 UCLA 活动评分。所有患者均在 1 年后进行随访。
两组患者的活动评分相似。小直径和大直径股骨头患者之间的金属离子水平或氧化应激标志物无差异。髋臼倾斜度和前倾角对金属离子水平没有影响。
数据表明,大直径或小直径金属对金属 THA 患者的离子值没有差异,增加的金属离子不会改变患者的氧化状态。