金属对金属髋关节表面置换术后,髋臼杯倾斜角度大于50度会增加钴和铬离子的全血浓度。

Cup inclination angle of greater than 50 degrees increases whole blood concentrations of cobalt and chromium ions after metal-on-metal hip resurfacing.

作者信息

Hart A J, Buddhdev P, Winship P, Faria N, Powell J J, Skinner J A

机构信息

Department of Musculoskeletal Surgery, Imperial College Healthcare NHS Trust, Charing Cross Hospital, London, UK.

出版信息

Hip Int. 2008 Jul-Sep;18(3):212-9. doi: 10.1177/112070000801800304.

Abstract

A cup inclination angle greater than 45 degrees is associated with increased wear rates of metal on polyethylene (MOP) hip replacements. The same maybe true for metal on metal (MOM) hips yet this has not been clearly shown. We measured the acetabular inclination angle from plain radiographs, and whole blood metal ion levels using Inductively Coupled Plasma Mass Spectrometry of 26 patients (mean Harris Hip Score 94 and mean time post op of 22 months) with Birmingham Hip Resurfacings. We identified a threshold level of 50 degrees cup inclination. Below this threshold, the mean whole blood cobalt and chromium were 1.6 ppb and 1.88 ppb respectively; above this threshold, the mean blood cobalt and chromium were 4.45 ppb and 4.3 ppb respectively. These differences were significant cobalt (p<0.01) and chromium (p=0.01). All patients above the threshold had metal levels greater than any of the patients below the threshold. For 14 patients, who returned one year later for a repeat blood metal level measurement, cobalt and chromium levels were very similar. The effect of an acetabular inclination angle of greater than 50 degrees on wear rates of MOM hips, as measured through blood metal ion levels, appears to be similar to that seen with MOP hips. Additionally, our new analytical methods may allow blood metal levels to be used as a realistic biomarker of in vivo wear rate of MOM hips. The implication is that metal levels can be minimised with optimal orientation of the acetabular component.

摘要

髋臼倾斜角度大于45度与金属对聚乙烯(MOP)髋关节置换物的磨损率增加有关。金属对金属(MOM)髋关节可能也是如此,但尚未得到明确证实。我们通过X线平片测量了26例行伯明翰髋关节表面置换术患者(平均Harris髋关节评分94分,平均术后时间22个月)的髋臼倾斜角度,并使用电感耦合等离子体质谱法测量了全血金属离子水平。我们确定了髋臼倾斜角度的阈值为50度。低于该阈值时,全血钴和铬的平均水平分别为1.6 ppb和1.88 ppb;高于该阈值时,全血钴和铬的平均水平分别为4.45 ppb和4.3 ppb。这些差异在钴(p<0.01)和铬(p=0.01)方面具有统计学意义。所有高于阈值的患者的金属水平均高于任何低于阈值的患者。对于14例一年后返回进行重复血金属水平测量的患者,钴和铬水平非常相似。通过血金属离子水平测量,髋臼倾斜角度大于50度对MOM髋关节磨损率的影响似乎与MOP髋关节相似。此外,我们的新分析方法可能使血金属水平用作MOM髋关节体内磨损率的实际生物标志物。这意味着通过优化髋臼组件的方向可以使金属水平降至最低。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验