Suppr超能文献

Development and characterization of sub-100 ps photomultiplier tubes.

作者信息

Horsfield C J, Rubery M S, Mack J M, Young C S, Herrmann H W, Caldwell S E, Evans S C, Sedilleo T J, Kim Y H, McEvoy A, Milnes J S, Howorth J, Davis B, O'Gara P M, Garza I, Miller E K, Stoeffl W, Ali Z

机构信息

AWE, Aldermaston, Reading RGR 4PR, United Kingdom.

出版信息

Rev Sci Instrum. 2010 Oct;81(10):10D318. doi: 10.1063/1.3475718.

Abstract

We describe the evaluation of a microchannel plate (MCP) photomultiplier tube (PMT), incorporating a 3 μm pore MCP and constant voltage anode and cathode gaps. The use of the small pore size results in PMTs with response functions of the order of 85 ps full-width-half-maximum, while the constant electric field across the anode and cathode gaps produces a uniform response function over the entire operating range of the device. The PMT was characterized on a number of facilities and employed on gas Cherenkov detectors fielded on various deuterium tritium fuel (DT) implosions on the Omega Laser Facility at the University of Rochester. The Cherenkov detectors are part of diagnostic development to measure Gamma ray reaction history for DT implosions on the National Ignition Facility.