Impurity temperature correction factors for the transmission grating spectrometer in the TJ-II stellarator.

作者信息

Arévalo J, McCarthy K J, Carmona J M, Fontdecaba J M

机构信息

Laboratorio Nacional de Fusión, Association Euratom-Ciemat, E-28040 Madrid, Spain.

出版信息

Rev Sci Instrum. 2010 Oct;81(10):10D705. doi: 10.1063/1.3475375.

Abstract

Impurity ion temperature and velocity profiles are obtained across plasmas in the TJ-II stellarator by performing charge-exchange recombination spectroscopy with a diagnostic neutral beam injector. For this, a tridirectional (toroidal plus two poloidal opposing views) multichannel spectroscopic diagnostic, incorporating 12-way fiber arrays, a compact f/1.8 spectrograph, and a back-illuminated CCD, permits Doppler line shifts and widths (of the C VI line at 529.05 nm) to be determined with 1-2 cm spatial resolution. For good photon counting statistics under Li-coated wall conditions, 600 μm diameter fibers collect and transmit light to curved 100 μm wide input slits. When calibrated with a neon pencil lamp this entrance slit width results in a non-Gaussian instrumental function that, if not handled correctly, can result in systematically underestimated impurity temperatures. Here we develop and present correction factors for this effect for a range of conditions.

摘要

相似文献

3
Toroidal charge exchange recombination spectroscopy measurements on MST.
Rev Sci Instrum. 2010 Oct;81(10):10D716. doi: 10.1063/1.3478665.
6
Wide-view charge exchange recombination spectroscopy diagnostic for Alcator C-Mod.
Rev Sci Instrum. 2008 Oct;79(10):10F529. doi: 10.1063/1.2979865.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验