Cone K V, Dunn J, Schneider M B, Baldis H A, Brown G V, Emig J, James D L, May M J, Park J, Shepherd R, Widmann K
Lawrence Livermore National Laboratory, Livermore, California 94551, USA.
Rev Sci Instrum. 2010 Oct;81(10):10E318. doi: 10.1063/1.3492407.
A 2400 lines/mm variable-spaced grating spectrometer has been used to measure soft x-ray emission (8-22 Å) from laser-produced plasma experiments at Lawrence Livermore National Laboratory's Compact Multipulse Terrawatt (COMET) Laser Facility. The spectrometer was coupled to a Kentech x-ray streak camera to study the temporal evolution of soft x rays emitted from the back of the Mylar and the copper foils irradiated at 10(15) W/cm(2). The instrument demonstrated a resolving power of ∼120 at 19 Å with a time resolution of 31 ps. The time-resolved copper emission spectrum was consistent with a photodiode monitoring the laser temporal pulse shape and indicated that the soft x-ray emission follows the laser heating of the target. The time and spectral resolutions of this diagnostic make it useful for studies of high temperature plasmas.
一台2400线/毫米可变间距光栅光谱仪被用于测量劳伦斯利弗莫尔国家实验室紧凑型多脉冲太瓦(COMET)激光装置上激光产生等离子体实验的软X射线发射(8 - 22埃)。该光谱仪与一台肯泰克X射线条纹相机耦合,以研究在10¹⁵瓦/平方厘米的辐照强度下,从聚酯薄膜和铜箔背面发射的软X射线的时间演化。该仪器在19埃处的分辨能力约为120,时间分辨率为31皮秒。时间分辨的铜发射光谱与监测激光时间脉冲形状的光电二极管一致,表明软X射线发射跟随靶材的激光加热。这种诊断方法的时间和光谱分辨率使其对高温等离子体的研究很有用。