Department of Medical Technology, Fooyin University, 151 Ching-Hsueh Road, Ta-Liao Hsiang, Kaohsiung Hsien, 831-02 Taiwan.
Analyst. 2011 Jan 21;136(2):393-400. doi: 10.1039/c0an00315h. Epub 2010 Nov 3.
It is a common knowledge that detector fatigue causes a calibration curve to deviate from the preferred linear relationship at the higher concentration end. With the adaptation of an isotopically labeled analog of the analyte as the internal standard (IS), cross-contribution (CC) of the intensities monitored for the ions designating the analyte and the IS can also result in a non-linear relationship at both ends. A novel approach developed to assess 'the extent and the effect of [CC]… in quantitative GC-MS analysis' can be extended (a) to examine whether a specific set of CC values is accurate; and (b) to differentiate whether the observed non-linear calibration curve is caused by detector fatigue or the CC phenomenon. Data derived from the exemplar secobarbital (SB)/SB-d(5) system (as di-butyl-derivatives) are used to illustrate this novel approach. Comparing the non-linear nature of calibration data that are empirically observed to that derived from theoretical calculation (with the incorporation of adjustment resulting from the ion CC phenomenon), supports the conclusions that (a) both CC and detector fatigue contribute significantly to the observed non-linear nature of the calibration curve based on ion-pair m/z 207/212; and (b) detector fatigue is the dominating contributor when the calibration curve is based on ion-pair m/z 263/268.
众所周知,检测器疲劳会导致校准曲线在较高浓度端偏离理想的线性关系。通过将分析物的同位素标记类似物作为内标(IS),用于标记分析物和 IS 的离子强度的交叉贡献(CC)也会导致在两端产生非线性关系。为评估“[CC]……在定量 GC-MS 分析中的程度和影响”而开发的一种新方法可以扩展(a)以检查一组特定的 CC 值是否准确;(b)区分观察到的非线性校准曲线是由检测器疲劳还是 CC 现象引起的。从示例仲戊巴比妥(SB)/SB-d(5)系统(作为二丁基衍生物)获得的数据用于说明这种新方法。将经验观察到的校准数据的非线性性质与理论计算得出的非线性性质(结合由于离子 CC 现象引起的调整)进行比较,支持以下结论:(a)CC 和检测器疲劳都会显著导致基于离子对 m/z 207/212 的校准曲线的非线性性质;(b)当校准曲线基于离子对 m/z 263/268 时,检测器疲劳是主要贡献者。