An ankle-foot orthosis with an oil damper (AFO-OD) was developed to resist plantarflexion motion, thereby improving hemiplegic gait performance. The purpose of this study was to determine the effect of AFO-OD on muscle activity during the gait cycle in individuals affected by stroke.
Electromyography (EMG) was used to assess gait at a self-selected speed while wearing an AFO-OD or an AFO with a plantarflexion stop (AFO-PS) worn on the affected side in 11 stroke survivors and on the right side in 11 age-matched healthy adults. EMG signals were obtained from the tibialis anterior (TA), gastrocnemius (GAS), and soleus (SOL) muscles. In addition, the ankle joint angle under both braces and the plantarflexion resistance torque (PFRT) under AFO-OD were monitored.
Peak PFRT under AFO-OD was observed during the loading response phase (LRP) in both groups. AFO-OD promoted adequate plantarflexion during LRP in the stroke group, whereas AFO-PS did not. Compared with the AFO-PS, the AFO-OD significantly reduced GAS EMG amplitude during LRP in the stroke group, which was significantly correlated with peak PFRT during LRP.
AFO-OD assisted the "heel rocker function" and reduced GAS muscle EMG amplitude during LRP.
一种带有油压阻尼器的踝足矫形器(AFO-OD)被开发出来以抵抗跖屈运动,从而改善偏瘫步态性能。本研究的目的是确定 AFO-OD 对脑卒中患者步态周期中肌肉活动的影响。
11 名脑卒中幸存者在患侧穿戴 AFO-OD,11 名年龄匹配的健康成年人在右侧穿戴带有跖屈限位装置的 AFO(AFO-PS),以自身选择的速度进行步态的肌电图(EMG)评估。从胫骨前肌(TA)、腓肠肌(GAS)和比目鱼肌(SOL)中获取 EMG 信号。此外,监测两种矫形器下的踝关节角度和 AFO-OD 下的跖屈阻力扭矩(PFRT)。
在两组中,AFO-OD 在负荷响应阶段(LRP)观察到峰值 PFRT。AFO-OD 促进了脑卒中患者 LRP 期间的充分跖屈,而 AFO-PS 则没有。与 AFO-PS 相比,AFO-OD 显著降低了脑卒中患者 LRP 期间的 GAS EMG 幅度,与 LRP 期间的峰值 PFRT 显著相关。
AFO-OD 辅助了“脚跟摆动功能”,并降低了 LRP 期间 GAS 肌肉的 EMG 幅度。