Svensson K L, Henriksson B A, Sonander H, Stenqvist O
Department of Anaesthesia and Intensive Care, Sahlgren's Hospital, Göteborg, Sweden.
Br J Anaesth. 1990 Mar;64(3):320-6. doi: 10.1093/bja/64.3.320.
A system for metabolic gas exchange has been used during nitrous oxide-opioid anaesthesia incorporating a Servo Ventilator 900 C and external analysers for oxygen and carbon dioxide. Oxygen consumption and carbon dioxide excretion were calculated as differences in content between inspired and expired minute ventilation. Nitrous oxide uptake was calculated similarly, assuming it was the only other gas present in addition to oxygen and carbon dioxide. The mean value for oxygen consumption was 3.25 ml kg-1 min-1, declining by 8% during the 2 h of anaesthesia. The formula for the best fit curve of nitrous oxide uptake was 18.3 . t-0.48 ml kg-1 min-1 when FIN2O was 0.7. To simplify measurement procedures and avoid measurements of expiratory volume, we also calculated metabolic gas exchange when expiratory minute ventilation was expressed as a function of inspiratory minute volume and nitrous oxide uptake. The latter value was obtained from the overall best fit curve for nitrous oxide uptake.
在氧化亚氮-阿片类麻醉期间,使用了一种代谢气体交换系统,该系统包含一台Servo Ventilator 900 C呼吸机以及用于氧气和二氧化碳的外部分析仪。氧气消耗量和二氧化碳排出量通过吸入和呼出分钟通气量之间的含量差异来计算。氧化亚氮摄取量的计算方式类似,假定它是除氧气和二氧化碳之外唯一存在的其他气体。氧气消耗量的平均值为3.25毫升·千克⁻¹·分钟⁻¹,在麻醉的2小时内下降了8%。当氧化亚氮终末吸入浓度(FIN2O)为0.7时,氧化亚氮摄取量的最佳拟合曲线公式为18.3·t⁻⁰.⁴⁸毫升·千克⁻¹·分钟⁻¹。为简化测量程序并避免测量呼出气量,当将呼气分钟通气量表示为吸气分钟量和氧化亚氮摄取量的函数时,我们还计算了代谢气体交换。后者的值从氧化亚氮摄取量的总体最佳拟合曲线中获得。