Nakajima Kazuki, Sekine Katsuhisa, Yamazaki Katsuya, Tampo Atsushi, Omote Yuuichirou, Fukunaga Hiroshi, Yagi Yasuko, Ishizu Kyoji, Nakajima Masanori, Tobe Kazuyuki, Kobayashi Masashi, Sasaki Kazuo
Division of Bio-Information Engineering, Faculty of Engineering, University of Toyama, 3190 Gofuku, Toyama, 930-8555, Japan.
Annu Int Conf IEEE Eng Med Biol Soc. 2010;2010:911-4. doi: 10.1109/IEMBS.2010.5627480.
The aim of this study is to develop a method for measuring the respiratory waveform using non-contact electrodes during bathing. To determine the most appropriate electrode arrangement, we modeled a composite system consisting of a body submerged in bath water. We calculated the frequency dependence of the impedance amplitude using a three-dimensional finite difference method (3D-FDM). The simulation results showed that an increase in chest size due to inspiration caused a decrease in the impedance amplitude in the frequency range of 0.1 Hz to 1 MHz. Next, bioelectric impedance (BEI) was measured in the frequency range of 4 kHz to 4 MHz at the maximum-end-expiration and maximum-end-inspiration stages. BEI results were consistent with those obtained from the model simulations. We found that 1 MHz was the appropriate frequency for measuring the respiratory waveform, and the time dependence of the impedance amplitude was measured at 1 MHz. The impedance amplitude agreed well with the respiratory waveform obtained from rubber strain gauge plethysmography, which was used as a reference.