College of Chemistry, Liaoning University, Shenyang, Liaoning 110036, PR China.
Talanta. 2010 Dec 15;83(2):332-6. doi: 10.1016/j.talanta.2010.09.024. Epub 2010 Sep 25.
A bismuth-film modified graphite nanofibers-Nafion glassy carbon electrode (BiF/GNFs-NA/GCE) was constructed for the simultaneous determination of trace Cd(II) and Pb(II). The electrochemical properties and applications of the modified electrode were studied. Operational parameters such as deposition potential, deposition time, and bismuth ion concentration were optimized for the purpose of determination of trace metal ions in 0.10 M acetate buffer solution (pH 4.5). Under optimal conditions, based on three times the standard deviation of the baseline, the limits of detection were 0.09 μg L(-1) for Cd(II) and 0.02 μg L(-1) for Pb(II) with a 10 min preconcentration. In addition, the BiF/GNFs-NA/GCE displayed good reproducibility and selectivity, making it suitable for the simultaneous determination of Cd(II) and Pb(II) in real sample such as river water and human blood samples.
构建了铋膜修饰石墨纳米纤维-纳滤玻璃碳电极(BiF/GNFs-NA/GCE),用于痕量 Cd(II) 和 Pb(II) 的同时测定。研究了修饰电极的电化学性质和应用。优化了沉积电位、沉积时间和铋离子浓度等操作参数,以在 0.10 M 乙酸盐缓冲溶液(pH 4.5)中测定痕量金属离子。在最佳条件下,基于基线的 3 倍标准偏差,Cd(II) 的检测限为 0.09 μg L(-1),Pb(II) 的检测限为 0.02 μg L(-1),预富集时间为 10 min。此外,BiF/GNFs-NA/GCE 显示出良好的重现性和选择性,适用于河水和人血等实际样品中 Cd(II) 和 Pb(II) 的同时测定。