Krasny R, Besgen J H
Klinik für Radiologische Diagnostik, Klinikum der RWTH Aachen.
Biomed Tech (Berl). 1990 Apr;35(4):72-7.
Gas have long been used in radiological examinations, in particular for the diagnostic evaluation of gastrointestinal diseases. However, not until the introduction of digital subtraction angiography was carbon dioxide also used with success for angiography of arteries. In order to advice exact and reproducible dosage of the gas with freely selectable injection parameters, a carbon dioxide gas metering device was developed. The principle of gas metering is based on the measurement of the pressure gradient of the gas across a defined pneumatic resistance. The gas metering devices was tested in 15 animal experiments to produce survey and selective angiograms.
气体长期以来一直用于放射学检查,特别是用于胃肠道疾病的诊断评估。然而,直到数字减影血管造影术问世,二氧化碳才成功用于动脉血管造影。为了在可自由选择注射参数的情况下精确且可重复地计量气体剂量,开发了一种二氧化碳气体计量装置。气体计量的原理基于测量气体在规定气动阻力上的压力梯度。该气体计量装置在15次动物实验中进行了测试,以制作概况和选择性血管造影照片。