Weissman Neil J, Turco Mark A, Ormiston John A, Mann Tift, Cannon Louis A, McGarry Thomas F, Webster Mark W I, Hall Jack J, Lucca Michael J, Mishkel Gregory J, Wu Chiung-Jen, Mandinov Lazar, Dawkins Keith D
MedStar Health Research Institute, Washington Hospital Center, Washington, DC 20010, USA.
Cardiovasc Revasc Med. 2011 Jul-Aug;12(4):247-57. doi: 10.1016/j.carrev.2010.09.008. Epub 2010 Dec 3.
The objective of this intravascular ultrasound (IVUS) analysis was to evaluate the vascular response of the thin-strut TAXUS Liberté stent compared with the otherwise identical TAXUS Express stent.
The TAXUS ATLAS and TAXUS ATLAS Long Lesion studies are nonrandomized trials comparing the thin-strut TAXUS Liberté stent to historical TAXUS Express controls from the TAXUS IV and TAXUS V trials. A total of 377 patients enrolled in the two TAXUS ATLAS studies were randomly selected for the IVUS subset and compared to 314 TAXUS Express IVUS controls.
Despite increased lesion complexity in the TAXUS Liberté group, neointimal formation at 9 months was similar in both stents (TAXUS Liberté 13.8±11.0%; TAXUS Express 13.1±13.8%, P=.56). However, this neointima covered more of the overall stent in the TAXUS Liberté (67.9±32.5%) compared with the TAXUS Express (54.4±37.2%, P<.001), suggesting more uniform neointimal distribution. TAXUS Liberté also showed less pronounced negative remodeling at both stent edges and had significantly less (4.3% vs. 9.6%, P=.04) late incomplete stent apposition (ISA).
Despite identical polymer and drug release characteristic, the thin-strut TAXUS Liberté stent demonstrates improved neointimal coverage, better edge remodeling, and less late ISA vs. TAXUS Express, hereby highlighting the importance of the platform design for drug-eluting stents.
本血管内超声(IVUS)分析的目的是评估薄支柱TAXUS Liberté支架与其他方面相同的TAXUS Express支架相比的血管反应。
TAXUS ATLAS和TAXUS ATLAS长病变研究是非随机试验,将薄支柱TAXUS Liberté支架与TAXUS IV和TAXUS V试验中的历史TAXUS Express对照进行比较。从两项TAXUS ATLAS研究中总共随机选择377例患者作为IVUS亚组,并与314例TAXUS Express IVUS对照进行比较。
尽管TAXUS Liberté组病变复杂性增加,但两种支架在9个月时的新生内膜形成相似(TAXUS Liberté为13.8±11.0%;TAXUS Express为13.1±13.8%,P = 0.56)。然而,与TAXUS Express(54.4±37.2%,P<0.001)相比,TAXUS Liberté中的这种新生内膜覆盖了更多的整个支架,表明新生内膜分布更均匀。TAXUS Liberté在两个支架边缘的负性重塑也不那么明显,晚期支架贴壁不全(ISA)明显更少(4.3%对9.6%,P = 0.04)。
尽管聚合物和药物释放特性相同,但与TAXUS Express相比,薄支柱TAXUS Liberté支架显示出更好的新生内膜覆盖、更好的边缘重塑和更少的晚期ISA,从而突出了药物洗脱支架平台设计的重要性。