Wang X F, Wang J E, Cao L S, Huang Y Z, Huang H Q, Deng Y B, Wu Y
Echocardiographic Laboratory, Union Hospital, Tongji Medical University, Wuhan.
Chin Med J (Engl). 1990 Feb;103(2):117-24.
Two-dimensional echocardiography was used to guide and locate the balloon of Swan-Ganz catheter during catheterization in 43 patients. The echoes of balloon appeared as a series of strong light spots, with a rate of detection of 100%. Under monitoring of two-dimensional echocardiography, the site of balloon and its moving route could be determined. The coincidence of echographic findings and X-ray findings showed that it can be used instead of X-ray monitoring to lessen irradiation during catheterization. The depth of catheter inserted was measured. The differential points between the echoes of balloon and tubing, the terminology of the balloon site in echocardiography and fluoroscopy, its clinical value and limitation are discussed.
在43例患者的导管插入术中,使用二维超声心动图来引导和定位Swan-Ganz导管的球囊。球囊的回声表现为一系列强光点,检测率为100%。在二维超声心动图监测下,可确定球囊的位置及其移动路径。超声检查结果与X线检查结果的一致性表明,它可用于替代X线监测,以减少导管插入术中的辐射。测量了导管插入的深度。讨论了球囊与导管回声的区别点、超声心动图和荧光透视中球囊位置的术语、其临床价值及局限性。