胎儿血管磁共振血管成像:在绵羊胎儿中的可行性研究。

Magnetic resonance angiography of fetal vessels: feasibility study in the sheep fetus.

机构信息

Department of Diagnostic and Interventional Radiology, University Medical Center Hamburg-Eppendorf, 52 Martinistrasse, Hamburg, 20246, Germany.

出版信息

Jpn J Radiol. 2010 Dec;28(10):720-6. doi: 10.1007/s11604-010-0498-x. Epub 2010 Dec 30.

Abstract

PURPOSE

The aim of this study was to perform fetal magnetic resonance angiography (MRA) in utero in a sheep model.

MATERIAL AND METHODS

Images of the great vessels, the heart, and the tracheal tree were performed on four pregnant ewes with a 1.5-T scanner (Philips Medical Systems, Best, The Netherlands). MRA was achieved in utero using a nontriggered free-breathing three-dimensional balanced fast field echo (FFE) technique. All obtained MRA images were evaluated in consensus on a three-point scale by two radiologists with 9 and 4 years of experience in fetal MRI, respectively.

RESULTS

The fetal heart frequencies were between 130 and 160 bpm. The aorta from the aortic bulb to the bifurcation as well as some of the main aortic branches could be depicted. The pulmonary trunk and arteries, the superior and inferior caval veins, and the subsegmental branches of the trachea could also be visualized.

CONCLUSION

The nontriggered MRA of the fetal great vessels with images of the tracheal tree allowed an excellent evaluation of anatomical structures.

摘要

目的

本研究旨在对绵羊模型进行宫内胎儿磁共振血管造影(MRA)。

材料与方法

使用 1.5T 扫描仪(荷兰飞利浦医疗系统公司)对 4 只怀孕的母羊进行大血管、心脏和气管树的图像检查。宫内 MRA 采用非触发自由呼吸三维平衡快速场回波(FFE)技术。由两位分别具有 9 年和 4 年胎儿 MRI 经验的放射科医生对所有获得的 MRA 图像进行共识评估,分别按照三点量表进行评估。

结果

胎儿心率在 130 到 160 bpm 之间。可以描绘从主动脉球囊到分叉的主动脉以及一些主要的主动脉分支。还可以看到肺动脉干和动脉、上腔静脉和下腔静脉以及气管的亚段分支。

结论

非触发式胎儿大血管 MRA 联合气管树图像可对解剖结构进行出色的评估。