Meredith K S, Monaco F J
Department of Neonatology, Memorial Hospital, Colorado Springs, Colorado 80909.
Pediatr Pulmonol. 1990;9(4):254-9. doi: 10.1002/ppul.1950090412.
We evaluated a new lightweight capnometer with a less than 1 mL deadspace neonatal airway adapter and endotracheal tube connector unit (NAC) for use in mechanically ventilated neonates. The evaluation consisted of: 1) a bench test comparison of air flow resistance between the standard endotracheal tube and connector with the new NAC (flow rates, 1.5 to 12.8 L/min); 2) a determination of the effect of NAC placement on Paco2; 3) pre- and post-NAC pulmonary mechanics; and 4) analysis of paired PetCO2 and PACO2 in 16 infants requiring mechanical ventilation. Paired t test of the slopes of the resistance curves was significant (P = 0.002) while analysis of variance of differential pressures was not (P = 0.29). All post-NAC placement Paco2 were smaller than pre-placement values; there were no differences in pulmonary mechanics, and Petco2 correlated closely with Paco2 (n = 132, r = 0.79) defined as Petco2 = 0.68.Paco2 + 5.52; means +/- 1 SD, Paco2 -Petco2 was 4.7 +/- 4.7 torr and Petco2/Paco2 was 0.86 +/- 0.14.
我们评估了一种新型轻便式二氧化碳监测仪,其配备了死腔小于1 mL的新生儿气道适配器和气管内导管连接单元(NAC),用于机械通气的新生儿。评估内容包括:1)在流量为1.5至12.8 L/分钟的情况下,对标准气管内导管及连接器与新型NAC之间的气流阻力进行台架试验比较;2)确定NAC放置对动脉血二氧化碳分压(Paco2)的影响;3)放置NAC前后的肺力学情况;4)对16例需要机械通气的婴儿的呼气末二氧化碳分压(PetCO2)和动脉血二氧化碳分压(PACO2)进行配对分析。阻力曲线斜率的配对t检验具有显著性(P = 0.002),而压差的方差分析则无显著性(P = 0.29)。放置NAC后所有的Paco2均小于放置前的值;肺力学情况无差异,且Petco2与Paco2密切相关(n = 132,r = 0.79),定义为Petco2 = 0.68·Paco2 + 5.52;均值±1标准差,Paco2 - Petco2为4.7±4.7托,Petco2/Paco2为0.86±0.14。