Suppr超能文献

可视喉镜引导下经口气管插管在模拟困难气道中的应用:管芯与普通喉镜的比较。

GlideScope Videolaryngoscopy in the Simulated Difficult Airway: Bougie vs Standard Stylet.

机构信息

University of California, Davis School of Medicine.

出版信息

West J Emerg Med. 2010 Dec;11(5):426-31.

Abstract

OBJECTIVE

GlideScope(®) videolaryngoscopy (GVL) has been shown to improve visualization of the glottis compared to direct laryngoscopy (DL). However, due to the angle of approach to the glottis, intubation can still be challenging. We hypothesized that novice GVL users would be able to intubate faster and easier using an airway introducer (frequently known as a bougie) than with a standard intubating stylet.

METHODS

Intubations were performed on a human airway simulator with settings for easy and difficult airways. Participants were emergency medicine (EM) residents or faculty (n=21) who were novice GVL users. Participants were intubated a total of eight times (four GVL, four DL) using either a bougie or an intubating stylet. We recorded time to intubate (TTI) and difficulty rating using a visual analog scale (VAS) and non-parametric statistical methods for analysis. We reported medians with interquartile range (IQR).

RESULTS

The median TTI with difficult airway settings and the bougie-GVL was 76 seconds (IQR 50, 102) versus 64 seconds (IQR 50.5, 125), p=0.76 for the stylet-GVL combination. The median VAS difficulty score, on difficult airway settings, for the bougie-GVL was 5 cm (IQR 3.3, 8.0) versus 6.2 cm (IQR 5.0, 7.5) with the stylet-GVL, p=0.53.

CONCLUSION

Among novices using GVL for simulated difficult airway management, there was no benefit, in terms of speed or ease of intubation, by using the bougie over the standard stylet.

摘要

目的

GlideScope(®)视频喉镜(GVL)与直接喉镜(DL)相比,可改善声门显露。然而,由于接近声门的角度,插管仍然具有挑战性。我们假设,与标准插管导丝相比,新手 GVL 用户使用气道导入器(通常称为引导管)进行插管会更快、更容易。

方法

在具有易于和困难气道设置的人体气道模拟器上进行插管。参与者为急诊医学(EM)住院医师或教员(n=21),他们是新手 GVL 用户。参与者总共使用引导管或插管导丝进行了 8 次插管(4 次 GVL,4 次 DL)。我们使用视觉模拟评分(VAS)记录插管时间(TTI)和难度评分,并使用非参数统计方法进行分析。我们报告中位数及其四分位距(IQR)。

结果

在困难气道设置下和使用引导管-GVL 时,TTI 的中位数为 76 秒(IQR 50,102),而使用插管导丝-GVL 时为 64 秒(IQR 50.5,125),p=0.76。在困难气道设置下,使用引导管-GVL 时,VAS 难度评分的中位数为 5 厘米(IQR 3.3,8.0),而使用插管导丝-GVL 时为 6.2 厘米(IQR 5.0,7.5),p=0.53。

结论

在使用 GVL 模拟困难气道管理的新手人群中,在速度或插管难易程度方面,使用引导管并没有优于标准插管导丝。