Suppr超能文献

丙氨酸/EPR 剂量测定在全身照射方案验证和使用人体模型进行治疗计划剂量计算中的应用。

Alanine/EPR dosimetry applied to the verification of a total body irradiation protocol and treatment planning dose calculation using a humanoid phantom.

机构信息

NuTeC-EPR Dosimetry Laboratory, Xios Hogeschool Limburg, Technologiecentrum 27, 3590 Diepenbeek, Belgium.

出版信息

Med Phys. 2010 Dec;37(12):6292-9. doi: 10.1118/1.3496355.

Abstract

PURPOSE

To avoid complications in total body irradiation (TBI), it is important to achieve a homogeneous dose distribution throughout the body and to deliver a correct dose to the lung which is an organ at risk. The purpose of this work was to validate the TBI dose protocol and to check the accuracy of the 3D dose calculations of the treatment planning system.

METHODS

Dosimetry based on alanine/electron paramagnetic resonance (EPR) was used to measure dose at numerous locations within an anthropomorphic phantom (Alderson) that was irradiated in a clinical TBI beam setup. The alanine EPR dosimetry system was calibrated against water calorimetry in a Co-60 beam and the absorbed dose was determined by the use of "dose-normalized amplitudes" A(D). The dose rate of the TBI beam was checked against a Farmer ionization chamber. The phantom measurements were compared to 3D dose calculations from a treatment planning system (Pinnacle) modeled for standard dose calculations.

RESULTS

Alanine dosimetry allowed accurate measurements which were in accordance with ionization chamber measurements. The combined relative standard measurement uncertainty in the Alderson phantom was U(r)(A(D))=0.6%. The humanoid phantom was irradiated to a reference dose of 10 Gy, limiting the lung dose to 7.5 Gy. The ratio of the average measured dose midplane in the craniocaudal direction to the reference dose was 1.001 with a spread of +/- 4.7% (1 sd). Dose to the lung was measured in 26 locations and found, in average, 1.8% lower than expected. Lung dose was homogeneous in the ventral-dorsal direction but a dose gradient of 0.10 Gy cm(-1) was observed in the craniocaudal direction midline within the lung lobe. 3D dose calculations (Pinnacle) were found, in average, 2% lower compared to dose measurements on the body axis and 3% lower for the lungs.

CONCLUSIONS

The alanine/EPR dosimetry system allowed accurate dose measurements which enabled the authors to validate their TBI dose protocol. Dose calculations based on a collapsed cone convolution dose algorithm modeled for regular treatments are accurate within 3% and can further be improved when the algorithm is modeled for TBI.

摘要

目的

为避免全身照射(TBI)中的并发症,确保全身剂量分布均匀并向肺(一种危险器官)提供正确剂量非常重要。本研究的目的是验证 TBI 剂量方案,并检查治疗计划系统的 3D 剂量计算的准确性。

方法

基于丙氨酸/电子顺磁共振(EPR)的剂量测量用于在一个仿体(Alderson)内的多个位置测量剂量,该仿体在临床 TBI 射束设置下进行照射。使用“剂量归一化幅度”A(D)将丙氨酸 EPR 剂量测量系统与 Co-60 射束中的水热量计进行校准,并使用“剂量归一化幅度”A(D)确定吸收剂量。使用 Farmer 电离室检查 TBI 射束的剂量率。将仿体测量值与为标准剂量计算建模的治疗计划系统(Pinnacle)的 3D 剂量计算进行比较。

结果

丙氨酸剂量测量允许进行准确测量,这些测量与电离室测量结果一致。在 Alderson 仿体中,综合相对标准测量不确定度 U(r)(A(D))为 0.6%。将类人仿体照射到参考剂量 10 Gy,将肺剂量限制在 7.5 Gy。颅尾方向的平均测量剂量与参考剂量的比值为 1.001,分布范围为 +/- 4.7%(1 sd)。在 26 个位置测量了肺剂量,平均比预期低 1.8%。肺剂量在腹背方向上是均匀的,但在肺叶中线的颅尾方向上观察到 0.10 Gy cm(-1)的剂量梯度。与身体轴上的剂量测量值相比,3D 剂量计算(Pinnacle)平均低 2%,与肺相比低 3%。

结论

丙氨酸/EPR 剂量测量系统允许进行准确的剂量测量,使作者能够验证他们的 TBI 剂量方案。基于为常规治疗建模的简化锥卷积剂量算法的剂量计算在 3%以内是准确的,并且当算法为 TBI 建模时可以进一步改进。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验