Clin Oral Implants Res. 2011 Oct;22(10):1179-1184. doi: 10.1111/j.1600-0501.2010.02089.x. Epub 2011 Feb 15.
The aim of this biomechanical study was to assess the influence of self-tapping blades in terms of primary implant stability between implants with self-tapping blades and implants without self-tapping blades using five different analytic methods, especially in medium-density bone.
Two different types of dental implants (4 × 10 mm) were tested: self-tapping and non-self-tapping. The fixture design including thread profiles was exactly the same between the two groups; the only difference was the presence of cutting blades on one half of the apical portion of the implant body. Solid rigid polyurethane blocks with corresponding densities were selected to simulate medium-density bone. Five mechanical assessments (insertion torque, resonance frequency analysis [RFA], reverse torque, pull-out and push in test) were performed for primary stability.
Implants without self-tapping blades showed significantly higher values (P<0.001) in four biomechanical assessments, except RFA (P=0.684). However, a statistically significant correlation could not be detected between insertion torque values with the four different outcome variables (P>0.05).
The outcomes of the present study indicate that the implant body design without self-tapping blades has a good primary stability compared with that with self-tapping blades in medium-density bone. Considering the RFA, a distinct layer of cortical bone on marginal bone will yield implant stability quotient values similar to those in medium-bone density when implants have the same diameter.
本生物力学研究旨在评估自攻刀片对种植体初期稳定性的影响,比较使用 5 种不同分析方法的自攻刀片种植体和非自攻刀片种植体之间的差异,尤其是在中密度骨中。
测试了两种不同类型的牙科种植体(4×10mm):自攻和非自攻。两组种植体的器械设计包括螺纹轮廓完全相同;唯一的区别是在种植体体部的根尖部分的一半上有切割刀片。选择具有相应密度的实心硬质聚氨酯块模拟中密度骨。进行了 5 项机械评估(插入扭矩、共振频率分析[RFA]、反向扭矩、拔出和推入试验)以评估初期稳定性。
除 RFA(P=0.684)外,在四项生物力学评估中,无自攻刀片的种植体显示出明显更高的数值(P<0.001)。然而,插入扭矩值与四个不同的结果变量之间未检测到统计学上的显著相关性(P>0.05)。
本研究结果表明,与自攻刀片种植体相比,在中密度骨中,无自攻刀片的种植体具有更好的初期稳定性。考虑到 RFA,当种植体具有相同直径时,边缘骨上明显的皮质骨层将产生类似于中密度骨的种植体稳定性商值。