Institut Fresnel, Aix-Marseille Université, CNRS, Ecole Centrale Marseille, Campus de Saint Jérôme, F-13397 Marseille cedex 20, France.
J Biomed Opt. 2011 Feb;16(2):021108. doi: 10.1117/1.3534783.
We perform multiplex coherent anti-Stokes Raman scattering (CARS) micro-spectroscopy with a picosecond pulsed laser and a broadband supercontinuum (SC) generated in photonic crystal fiber. CARS signal stability is achieved using an active fiber coupler that avoids thermal and mechanical drifts. We obtain multiplex CARS spectra for test liquids in the 600-2000 cm(-1) spectral range. In addition we investigate the polarization dependence of the CARS spectra when rotating the pump beam linear polarization state relative to the linearly polarized broad stokes SC. From these polarization measurements we deduce the Raman depolarization ratio, the resonant versus nonresonant contribution, the Raman resonance frequency, and the linewidth.
我们使用皮秒脉冲激光器和在光子晶体光纤中产生的宽带超连续谱进行了多路相干反斯托克斯拉曼散射(CARS)微光谱学实验。通过使用主动光纤耦合器实现了 CARS 信号稳定性,避免了热和机械漂移。我们获得了在 600-2000cm(-1)光谱范围内测试液体的多路 CARS 光谱。此外,我们还研究了在相对于线性偏振宽斯托克斯 SC 旋转泵浦光束线性偏振状态时 CARS 光谱的偏振依赖性。从这些偏振测量中,我们推导出了拉曼退偏振比、共振与非共振贡献、拉曼共振频率和线宽。