Institute of Diagnostic and Interventional Radiology, University Hospital, Friedrich Schiller University Jena, Germany.
Acad Radiol. 2011 May;18(5):634-9. doi: 10.1016/j.acra.2010.12.011. Epub 2011 Mar 2.
The aim of this study was to evaluate the suitability of 3-T magnetic resonance imaging (MRI) for olfactory bulb volumetry, comparing image quality obtained using different sequences on the basis of physical characteristics in combination with observer performance.
Twenty-two healthy volunteers (11 men, 11 women; mean age, 25 years) underwent 3-T MRI of the frontal skull base in this prospective study. Imaging was performed using a constructive interference in steady state (CISS) three-dimensional Fourier transformation sequence, a three-dimensional T2-weighted (3D-T2w) sequence, and a two-dimensional T2-weighted (2D-T2w) sequence. The relative performance of sequences was assessed using visual grading characteristic analysis. Interobserver agreement was assessed using κ statistics. For evaluation of physical image quality characteristics, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated and compared using Wilcoxon's test. SNR and CNR measurements were correlated with visual grading results.
Visual grading characteristic analysis showed significantly better image quality ratings for the CISS sequence compared to the 3D-T2w and 2D-T2w sequence, and the 2D-T2w sequence performed significantly better compared to the 3D-T2w sequence (P < .001). Interobserver agreement was substantial (κ = 0.66-0.73). Wilcoxon's test revealed significantly higher SNR and CNR values for the 2D-T2w sequence compared to the 3D-T2w and CISS sequences, and SNR and CNR values for the 3D-T2w sequence were significantly higher compared to those for the CISS sequence (P < .001 for each). Significant correlation between SNR and CNR and visual grading was found for the 3D-T2w sequence (SNR: ρ = 0.510, P = .015; CNR: ρ = 0.546, P = .009).
High-resolution 3-T MRI resulted in excellent values for SNR and CNR, suggesting the appropriateness of the sequences for olfactory bulb MRI volumetry. Visual grading characteristic analysis revealed the CISS sequence to be the most suitable for olfactory bulb volumetry.