Center for Non-Destructive Testing (RECENDT) GmbH, Linz, Austria. ro.nuster@uni‐graz.at
Opt Lett. 2011 Mar 15;36(6):981-3. doi: 10.1364/OL.36.000981.
An optical detection setup consisting of a focused laser beam fed into a downstream Fabry-Perot interferometer (FPI) for demodulation of acoustically generated optical phase variations is investigated for its applicability in photoacoustic tomography. The device measures the time derivative of acoustic signals integrated along the beam. Compared to a setup where the detection beam is part of a Mach-Zehnder interferometer, the signal-to-noise ratio of the FPI is lower, but the image quality of the two devices is similar. Using the FPI in a photoacoustic tomograph allows scanning the probe beam around the imaging object without moving the latter.
研究了一种光学检测装置,该装置由聚焦激光束进入下游法布里-珀罗干涉仪(FPI)组成,用于解调声致光学相位变化,以评估其在光声断层摄影中的适用性。该装置测量沿光束积分的声信号的时间导数。与检测光束是马赫-曾德尔干涉仪一部分的装置相比,FPI 的信噪比较低,但两种装置的图像质量相似。在光声断层扫描仪中使用 FPI 可以在不移动成像物体的情况下扫描探头光束。