Cardiology Department, National University Heart Centre Singapore, 1E Kent Ridge Road, NUHS Tower Block Level 9, Singapore 119228, Singapore.
Europace. 2011 Jun;13(6):825-8. doi: 10.1093/europace/eur071. Epub 2011 Mar 23.
Implantable cardioverter defibrillators (ICDs) reduce sudden death in patients at high risk. High defibrillation thresholds (DFTs) are not uncommon and may be the cause of failed defibrillation in patients with ICDs. Addition of a coil in the azygous vein posterior to the heart lowers the DFT in most patients by altering the electrical vector during defibrillation. This can be accomplished fairly easily and expeditiously using standard equipment in the laboratory. It also avoids the difficulties and complications associated with other methods such as the use of a subcutaneous array.
This series of three cases illustrates the type of patients who may benefit from this technique. The addition of an azygous coil successfully lowered the DFT to <10 J in each case. A brief description of the technique used to implant the azygous coil follows.
This simple and elegant solution can be highly useful as part of the armamentarium in tackling the problem of high DFTs.
植入式心脏复律除颤器(ICD)可降低高危患者的猝死风险。高除颤阈值(DFT)并不少见,可能是 ICD 患者除颤失败的原因。在心脏后方的奇静脉中增加一个线圈,通过在除颤过程中改变电矢量,可使大多数患者的 DFT 降低。这可以使用实验室中的标准设备相当容易和迅速地完成。它还避免了与其他方法(如使用皮下数组)相关的困难和并发症。
这一系列三个病例说明了可能从该技术中受益的患者类型。在每个病例中,添加奇静脉线圈都成功地将 DFT 降低至<10J。下面简要描述了植入奇静脉线圈所使用的技术。
作为解决高 DFT 问题的手段之一,这种简单而优雅的解决方案非常有用。