Yoo Y E, Lee K H, Je T J, Choi D S, Kim S K
Nano-Mechanical Systems Research Division, Korea Institute of Machinery and Materials, Daejeon 305-343, Korea.
J Nanosci Nanotechnol. 2011 Jan;11(1):233-8. doi: 10.1166/jnn.2011.3287.
Micro or nanofluidic devices have many channel shapes to deliver chemical solutions, body fluids or any fluids. The channels in these devices should be covered to prevent the fluids from overflowing or leaking. A typical method to fabricate an enclosed channel is to bond or weld a cover plate to a channel plate. This solid-to-solid bonding process, however, takes a considerable amount of time for mass production. In this study, a new process for molding a cover layer that can enclose open micro or nanochannels without solid-to-solid bonding is proposed and its feasibility is estimated. First, based on the design of a model microchannel, a brass microchannel master core was machined and a plastic microchannel platform was injection-molded. Using this molded platform, a series of experiments was performed for four process or mold design parameters. Some feasible conditions were successfully found to enclosed channels without filling the microchannels for the injection molding of a cover layer over the plastic microchannel platform. In addition, the bond strength and seal performance were estimated in a comparison with those done by conventional bonding or welding processes.
微流体或纳流体装置有多种通道形状,用于输送化学溶液、体液或任何流体。这些装置中的通道需要覆盖,以防止流体溢出或泄漏。制造封闭通道的一种典型方法是将盖板粘结或焊接到通道板上。然而,这种固态到固态的粘结过程在大规模生产中需要花费大量时间。在本研究中,提出了一种用于模制覆盖层的新工艺,该工艺可以在不进行固态到固态粘结的情况下封闭开放的微通道或纳通道,并评估了其可行性。首先,基于一个模型微通道的设计,加工了一个黄铜微通道母芯,并注塑成型了一个塑料微通道平台。利用这个成型平台,针对四个工艺或模具设计参数进行了一系列实验。成功找到了一些可行的条件,用于在不填充微通道的情况下封闭通道,以便在塑料微通道平台上注塑覆盖层。此外,还与传统粘结或焊接工艺进行了比较,评估了粘结强度和密封性能。