Suppr超能文献

Optimizing interoperability between video-oculographic and electromyographic systems.

作者信息

Navallas Javier, Ariz Mikel, Villanueva Arantxa, San Agustín Javier, Cabeza Rafael

机构信息

Department of Electrical and Electronic Engineering, Public University of Navarra, Campus de Arrosadia, Pamplona, Navarra, Spain.

出版信息

J Rehabil Res Dev. 2011;48(3):253-65. doi: 10.1682/jrrd.2010.06.0112.

Abstract

A new system is presented that enhances the interoperability between a video-oculographic (VOG) system for mouse movement control and an electromyographic (EMG) system for mouse click detection. The proposed VOG-EMG system combines gaze and muscle information to minimize the number of undesired clicks due to involuntary activations and environmental noise. We tested the system with 24 subjects, comparing three different configurations: one in which the VOG and EMG systems worked independently and two in which we used VOG gaze information to improve the EMG click detection. Results show that the number of false-positive click detections can be reduced when VOG and EMG information is combined. In addition, the third configuration, including extra processing, can reduce the activation delay produced because of the combined use of the VOG and EMG systems. The new VOG-EMG system is meant to be used in noisy environments in which the number of false clicks may impeach a reliable human-computer interaction.