Suppr超能文献

用于在表面活性剂存在下测定痕量金属的一次性 Nafion 修饰微制造铋膜传感器的溶出伏安分析。

Disposable Nafion-modified micro-fabricated bismuth-film sensors for voltammetric stripping analysis of trace metals in the presence of surfactants.

机构信息

Laboratory of Analytical Chemistry, Department of Chemistry, University of Athens, 157 71 Athens, Greece.

出版信息

Talanta. 2011 May 15;84(3):696-701. doi: 10.1016/j.talanta.2011.01.077. Epub 2011 Feb 4.

Abstract

This work is a study of the analytical utility of Nafion-modified microfabricated bismuth film electrodes (BiFEs) for the determination of Pb(II) and Cd(II) by anodic stripping voltammetry (ASV) in the presence of surfactants. Micro-fabricated BiFEs were prepared by depositing a thin film of bismuth on the surface of a silicon substrate by sputtering while the two-dimensional geometry of the final sensors was defined by photolithography. The BiFEs were further drop-coated with a Nafion film. These devices were applied to the determination of Pb(II) and Cd(II) by square wave ASV (SWASV) in the presence of Triton X-100 (a non-ionic surfactant), cetyltrimethylammonium bromide (CTAB) (a cationic surfactant) and sodium dodecyl sulphate (SDS) (an anionic surfactant). It was found that the presence of Nafion afforded an increase in sensitivity and the tolerance against surfactants but these properties were severely influenced by both the thickness of the Nafion film and the nature of the interfering surfactant. Using a Nafion of 0.4 μm thickness and 120 s of preconcentration, the repeatability (expressed as the % relative standard deviation on the same sensor (n=8)) at the 20 μg l(-1) level was 3.8% for Pb(II) and 3.1% for Cd(II) and the limits of detection were 0.5 μg l(-1) for Cd(II) and Pb(II). The sensors were applied to Cd(II) and Pb(II) determination in a certified lake-water sample.

摘要

这项工作研究了在表面活性剂存在下,通过阳极溶出伏安法(ASV)用 Nafion 修饰的微制造铋膜电极(BiFE)对 Pb(II)和 Cd(II)的分析应用。通过溅射在硅衬底表面沉积铋薄膜来制备微制造 BiFE,而最终传感器的二维几何形状则通过光刻定义。然后将 BiFE 进一步滴涂一层 Nafion 膜。这些器件用于在 Triton X-100(一种非离子表面活性剂)、十六烷基三甲基溴化铵(CTAB)(一种阳离子表面活性剂)和十二烷基硫酸钠(SDS)(一种阴离子表面活性剂)存在下通过方波阳极溶出伏安法(SWASV)测定 Pb(II)和 Cd(II)。结果发现,Nafion 的存在提高了灵敏度和对表面活性剂的耐受性,但这些特性受到 Nafion 膜的厚度和干扰表面活性剂的性质的严重影响。使用 0.4 μm 厚的 Nafion 和 120 s 的预浓缩,在 20 μg l(-1) 水平下,同一传感器(n=8)的重复性(表示为相对标准偏差的百分比)对于 Pb(II)为 3.8%,对于 Cd(II)为 3.1%,检测限分别为 0.5 μg l(-1)对于 Cd(II)和 Pb(II)。该传感器用于测定认证湖水样品中的 Cd(II)和 Pb(II)。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验