Oshagh Morteza, Hematian Mohammad Rahim, Shahidi Shoaleh, Feizi Negar, Bayani Farnaz, Pishbin Lida
Department of Orthodontics, School of Dentistry, Shiraz University of Medical Sciences, Shiraz, Iran.
World J Orthod. 2010 Winter;11(4):e72-7.
To assess whether the forces created by the magnetic resonance imaging (MRI) field cause a deflection of orthodontic stainless steel archwires or a change of their mechanical properties.
Sixty stainless steel 0.016-inch archwires were divided into three groups of 20 wires each. The control group was as received from the manufacturer, the MRI 1 group had wires attached to brackets on artificial teeth subjected to MRI (1.5 Tesla), and the MRI 2 group had activated T-loops subjected to MRI (1.5 Tesla). The mechanical properties were compared between the control group and MRI 1 group, whereas in the MRI 2 group, the change in the amount of activation was measured. The data were evaluated by repeated measurement analysis with independent and paired t tests.
Tensile strength did not differ significantly between the control group and the MRI 1 group (P=.64). In the MRI 2 group, the mean distance between the vertical loop legs before and after MRI did not differ significantly (P=.53).
The magnetic field did not significantly affect the mechanical properties of the tested stainless steel wires, nor did it change the loop activation.
评估磁共振成像(MRI)场产生的力是否会导致正畸不锈钢弓丝发生偏转或改变其机械性能。
将60根0.016英寸的不锈钢弓丝分为三组,每组20根。对照组为从制造商处收到的弓丝,MRI 1组的弓丝连接在接受MRI(1.5特斯拉)检查的人工牙齿上的托槽上,MRI 2组的激活T形曲接受MRI(1.5特斯拉)检查。比较对照组和MRI 1组之间的机械性能,而在MRI 2组中,测量激活量的变化。数据通过独立和配对t检验的重复测量分析进行评估。
对照组和MRI 1组之间的拉伸强度无显著差异(P = 0.64)。在MRI 2组中,MRI前后垂直曲腿之间的平均距离无显著差异(P = 0.53)。
磁场对测试的不锈钢丝的机械性能没有显著影响,也没有改变曲的激活状态。