Suppr超能文献

Viabahn覆膜支架移植物因血栓消融术导致的塌陷:AngioJet装置的一种罕见并发症。

Collapse of Viabahn stent-graft secondary to rheolytic thrombectomy: a rare complication of the AngioJet device.

作者信息

Tan Tze-Woei, Watanabe Meguru, Kabutey Nii-Kabu, Vilvendhan Rajendran, Kim Ducksoo

机构信息

Section of Vascular and Endovascular Surgery, Boston University Medical Center and Boston University School of Public Health, Boston, MA, USA.

出版信息

Ann Vasc Surg. 2011 May;25(4):557.e15-7. doi: 10.1016/j.avsg.2010.09.028.

Abstract

There is increased use of percutaneous mechanical thrombectomy for treatment of occluded dialysis access. The AngioJet rheolytic thrombectomy device is one such device available. Reports have shown safety and efficacy of these techniques with relatively few complications. We describe a case report of a collapsed Viabahn endoprosthesis in an arteriovenous fistula during treatment with an AngioJet device.

摘要

经皮机械血栓切除术在治疗闭塞性透析通路中的应用日益增加。AngioJet 流变血栓切除术装置就是其中一种可用的装置。报告显示这些技术具有安全性和有效性,并发症相对较少。我们描述了一例在使用 AngioJet 装置治疗过程中动静脉瘘内 Viabahn 内支架塌陷的病例报告。