Suppr超能文献

起搏器对 50 Hz 噪声电压的灵敏度。

Pacemaker sensitivity to 50 Hz noise voltages.

机构信息

Faculty of Medicine, Justus-Liebig-University, Friederichstr. 18, Giessen DE-35392, Germany.

出版信息

Europace. 2011 Sep;13(9):1319-26. doi: 10.1093/europace/eur121. Epub 2011 May 7.

Abstract

AIMS

European standards specify that pacemakers (PM) should be resistant to electromagnetic interference (EMI) up to an upper borderline voltage as a function of frequency. Electromagnetic interference fields should either remain below this upper borderline voltage level or be identified and isolated from the general population. Physicians caring for PM patients need to be aware of potential problems relating to EMI. For example, sensitivity should be programmed to avoid sensing EMI below the recommended borderline voltage level. The susceptibility of a pacemaker (PM) to 50 Hz noise is an important parameter of EMI and depends on the programmed sensitivity [sensitivity setting (SS)]. We studied SS and 50 Hz noise thresholds in a large population and determined the borderline SS, defined as the SS below which 50 Hz noise was sensed. Our results should be taken into consideration in programming the SS to protect patients from the adverse effects of EMI.

METHODS AND RESULTS

Measurements were performed on 189 PMs explanted after death. All PMs studied were implanted in 1998 or later. Sensitivity setting and sensing configuration (unipolar or bipolar) were left as programmed during lifetime. The ventricular SS and 50 Hz noise thresholds were measured according to the European pacemaker standard. The signal-to-noise ratios (S/N) were derived from the heart test and noise test signal thresholds. The S/N for pulsed 50 Hz noise of five manufacturers tested ranged from 0.435 to 0.59. The S/N for 50 Hz continuous noise for four manufactures other than Medtronic was higher, ranging from 0.458 to 0.623. No PM showed a ratio of 1 or better. Medtronic PMs reacted differently to 50 Hz continuous noise than the other brands. In 24 Medtronic PMs, the continuous noise threshold was evaluated with two heart test signal amplitudes: either 10 mV or threshold level. In tests at threshold amplitudes, voltages between 0.1 and 0.85 mV elicited interference proving that Medtronic PM reacted extremely sensitively to noise. At 10 mV heart test amplitude, the noise threshold was inversely proportional to the SS, i.e. higher SS resulted in lower noise thresholds. Noise immunity increases with increasing heart test signal amplitude.

CONCLUSION

All tested PMs reacted to pulsed 50 Hz waves as if they were heart signals and were inhibited. Continuous noise above noise threshold evoked asynchronous pacing at noise rate. All PMs had an S/N ratio <1, indicating that the heart signals were amplified less than noise. The European Standard requires that unipolarly sensing PMs tolerate noise up to 2 mV. However, an SS of 2 mV does not guarantee a noise tolerance of 2 mV. In order to fulfil this requirement, SS in the majority of PMs must be programmed >2 mV. In Medtronic PMs, the continuous noise threshold is paradoxical as it is higher with decreasing SS. As a good compromise in Medtronic PMs, SS should be ∼3 mV to guarantee sufficient protection from pulsed and continuous noise, assuming ventricular heart signals of 10 mV or more.

摘要

目的

欧洲标准规定,起搏器(PM)应能抵抗电磁干扰(EMI),最高边界电压与频率有关。电磁干扰场应低于此上限电压水平,或被识别并与普通人群隔离。照顾 PM 患者的医生需要了解与 EMI 相关的潜在问题。例如,应将灵敏度编程为避免在推荐的边界电压水平以下感应 EMI。起搏器对 50 Hz 噪声的敏感性是 EMI 的一个重要参数,取决于编程的灵敏度[灵敏度设置(SS)]。我们在大量人群中研究了 SS 和 50 Hz 噪声阈值,并确定了边界 SS,定义为低于该 SS 时可感知到 50 Hz 噪声。我们的研究结果应在编程 SS 时加以考虑,以保护患者免受 EMI 的不利影响。

方法和结果

对 189 例死后取出的 PM 进行了测量。研究的所有 PM 均于 1998 年或以后植入。在整个生命周期中,将 SS 和感应配置(单极或双极)保持为编程时的设置。根据欧洲起搏器标准测量心室 SS 和 50 Hz 噪声阈值。从心脏测试和噪声测试信号阈值得出信噪比(S/N)。五个制造商测试的脉冲 50 Hz 噪声的 S/N 范围为 0.435 至 0.59。四个制造商(除 Medtronic 外)的 50 Hz 连续噪声的 S/N 更高,范围为 0.458 至 0.623。没有 PM 的比值为 1 或更好。与其他品牌相比,Medtronic PM 对 50 Hz 连续噪声的反应不同。在 24 个 Medtronic PM 中,使用两种心脏测试信号幅度评估连续噪声阈值:10 mV 或阈值水平。在阈值幅度测试中,0.1 至 0.85 mV 的电压会引起干扰,证明 Medtronic PM 对噪声非常敏感。在 10 mV 的心脏测试幅度下,噪声阈值与 SS 成反比,即 SS 越高,噪声阈值越低。随着心脏测试信号幅度的增加,噪声免疫力增加。

结论

所有测试的 PM 对脉冲 50 Hz 波的反应都像是心脏信号并被抑制。高于噪声阈值的连续噪声会以噪声率引起异步起搏。所有 PM 的 S/N 比值均<1,表明心脏信号的放大程度低于噪声。欧洲标准要求单极感应 PM 能耐受高达 2 mV 的噪声。然而,2 mV 的 SS 并不能保证 2 mV 的噪声耐受。为了满足这一要求,大多数 PM 的 SS 必须编程为>2 mV。在 Medtronic PM 中,连续噪声阈值是矛盾的,因为随着 SS 的降低,噪声阈值会升高。作为 Medtronic PM 的一个很好的折衷方案,SS 应该约为 3 mV,以确保对脉冲和连续噪声有足够的保护,假设心室心脏信号为 10 mV 或更高。

相似文献

1
Pacemaker sensitivity to 50 Hz noise voltages.
Europace. 2011 Sep;13(9):1319-26. doi: 10.1093/europace/eur121. Epub 2011 May 7.
2
Do induction cooktops interfere with cardiac pacemakers?
Europace. 2006 May;8(5):377-84. doi: 10.1093/europace/eul014. Epub 2006 Mar 23.
3
Investigation of pacemaker position, lead configuration, and sensitivity setting in pacemakers of 579 deceased patients.
Europace. 2011 Jan;13(1):96-101. doi: 10.1093/europace/euq405. Epub 2010 Nov 17.
9
Interference by cellular phones with permanent implanted pacemakers: an update.
Europace. 2004 Jul;6(4):363-9. doi: 10.1016/j.eupc.2004.03.010.
10
Bench study of the accuracy of a commercial AED arrhythmia analysis algorithm in the presence of electromagnetic interferences.
Physiol Meas. 2009 Jul;30(7):695-705. doi: 10.1088/0967-3334/30/7/012. Epub 2009 Jun 12.

引用本文的文献

1
Adding wisdom to 'smart' bioelectronic systems: a design framework for physiologic control including practical examples.
Bioelectron Med (Lond). 2019 Mar;2(1):29-41. doi: 10.2217/bem-2019-0008. Epub 2019 May 30.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验