[20台起搏器在患者通过两种型号的防盗检测门时的表现]

[Behavior of 20 pacemakers as patients pass through 2 models of theft-detection doors].

作者信息

Frank R

机构信息

Centre de stimulation cardiaque, hôpital Jean-Rostand, 39-41, rue Jean-Le-Galleu, 94200 Ivry-sur-Seine, France.

出版信息

Ann Cardiol Angeiol (Paris). 2000 Jun;49(3):187-97.

Abstract

Episodic interference between the magnetic field of anti thief devices in stores (Electronic Article Surveillance, EAS) and pacemaker function have been reported, but no systematic in vivo study. 20 implanted pacemakers from ten different companies, 1 VVIR, 1 VDD and 18 DDD, were tested with two different EAS systems. #1 uses a 7.5 and 5 kHz field of 670 microTesla and #2 a 72.3 Hz magnetic field, of 1.10 microTesla. Under continuous ECG monitoring, patients had to walk through, and then stay between the doors while moving, and finally to lay against each wall. This was done with their usual pacemaker setting, and then set at their highest sensitivity in the unipolar atrial and ventricular mode. No abnormal PM behavior was observed with standard settings (V sensitivity 1.4 to 5 mV, A sensitivity 0.25 to 1 mV, bipolar (14) or unipolar (6). When programmed at the highest sensitivity, the magnetic field was detected by 5 models, but only with ventricular sensing set at less than 1.2 mV. moreover this only occurred in #2 EAS and not in #1. This resulted in short inhibitions < 1.2 s, and in one safety zone pacing. All occurred in devices conceived before 1993 (6/14) and none after (0/6). EAS devices should not inhibit implanted PM with standard settings, especially in recent models. Ventricular sensitivity should not be programmed in the unipolar mode at 1.2 mV or less.

摘要

已有报道称商店防盗设备(电子商品防盗系统,EAS)的磁场与起搏器功能之间存在偶发性干扰,但尚无系统的体内研究。对来自十家不同公司的20台植入式起搏器(1台VVIR、1台VDD和18台DDD)使用两种不同的EAS系统进行了测试。系统1使用670微特斯拉的7.5千赫兹和5千赫兹磁场,系统2使用1.10微特斯拉的72.3赫兹磁场。在持续心电图监测下,患者需走过门,然后在门之间移动时停留,最后靠在每面墙边。这一过程在患者平时的起搏器设置下进行,然后在单极心房和心室模式下将其设置为最高灵敏度。在标准设置下(V灵敏度为1.4至5毫伏,A灵敏度为0.25至1毫伏,双极(14台)或单极(6台))未观察到起搏器的异常行为。当设置为最高灵敏度时,有5款型号检测到了磁场,但仅在心室感知设置低于1.2毫伏时出现。此外,这种情况仅发生在系统2的EAS中,而未发生在系统1中。这导致了小于1.2秒的短暂抑制以及一次安全区起搏。所有情况均发生在1993年之前生产的设备中(6/14),1993年之后生产的设备中未出现(0/6)。EAS设备在标准设置下不应抑制植入式起搏器,尤其是在较新的型号中。心室灵敏度不应在单极模式下设置为1.2毫伏或更低。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验