Department of Palliative Medicine, Tan Tock Seng Hospital, 11 Jalan Tan Tock Seng, Singapore 308433.
Singapore Med J. 2011 Apr;52(4):e75-8.
Various laboratory and patient-related factors can result in falsely high or low glycated haemoglobin (HbA1c) measurements, and haemoglobin (Hb) variants that interfere with laboratory readings is an important cause of this. We report a case of a rare Hb variant, Hb Santa Juana, manifesting as a falsely high HbA1c in a 62-year-old patient with type 2 diabetes mellitus. The patient presented with high HbA1c values that persisted despite the intensification of anti-diabetic treatment. His home blood glucose levels were incongruently low compared to his HbA1c values. Further investigations revealed a family history of the variant Hb Santa Juana. This was confirmed in the patient when his blood was sent for DNA analysis. It is important for clinicians to be aware of the factors that can influence laboratory HbA1c measurements, as clinical decisions on treatment are often based on these measurements.
各种实验室和患者相关因素可导致糖化血红蛋白(HbA1c)检测值偏高或偏低,而干扰实验室检测结果的血红蛋白(Hb)变异是导致这种情况的一个重要原因。我们报告了 1 例罕见的 Hb 变异型,即 Hb Santa Juana,该变异型可导致 1 例 62 岁 2 型糖尿病患者的 HbA1c 检测值偏高。该患者的 HbA1c 值一直偏高,尽管已加强了降糖治疗,但该值仍居高不下。与 HbA1c 值相比,其家庭血糖值明显偏低。进一步的检查发现该患者存在变异型 Hb Santa Juana 的家族史。当患者的血液被送去进行 DNA 分析时,证实了这一点。临床医生了解可影响实验室 HbA1c 检测值的因素非常重要,因为治疗的临床决策通常基于这些检测值。