Suppr超能文献

经机械牵张成骨器的血流:初步试验。

Flow through a mechanical distraction enterogenesis device: a pilot test.

机构信息

Section of Pediatric Surgery, Department of Surgery and the C.S. Mott Children's Hospital, University of Michigan, Ann Arbor, Michigan, USA.

出版信息

J Surg Res. 2011 Oct;170(2):179-84. doi: 10.1016/j.jss.2011.03.058. Epub 2011 Apr 19.

Abstract

BACKGROUND

We tested the coupling portion of a prototype intraluminal distraction enterogenesis device to allow flow-through of simulated enteric contents (SEC) in both pig and human jejunum.

MATERIALS AND METHODS

SEC was made using 80% corn syrup. Ten-cm pig and human intestinal segments had a spoke-shaped 2.2 cm coupling adaptor sutured in place, intraluminally. The adaptor had a flow-through area of 33.6 mm(2). SEC was pumped into the proximal part of the intestinal segment at 0.083 mL/s. The times to first passage of SEC through the coupler (first drop), 10 mL, and 20 mL of SEC eluted from the distal end were recorded.

RESULTS

Mean time to first drop elution was 155 ± 38 s with pig, and 149 ± 22 s with human bowel (P = 0.8). This corresponded to a hydrostatic pressure of 37.5 mmHg before the initial drop passed through. Mean flow rates were 0.094 mL/s in pig bowel and 0.084 mL/s in human bowel (P = 0.09). To account for occlusion from luminal debris, a 75% occlusion of coupler holes was studied in the smaller pig bowel to investigate if reductions in flow-through area could be tolerated. Mean time to first drop increased slightly to 171 ± 15 s, but the elution rate stayed the same (P = 0.5).

CONCLUSIONS

After a physiologic level of initial pressure buildup allowing the first drop of SEC to pass the coupling adaptor, our prototype intestinal coupling adaptor did not obstruct flow-through of SEC, even after a 75% decrease in flow-through area. This type of attachment represents a viable approach to placing a device in-continuity without obstructing flow of enteric contents.

摘要

背景

我们测试了一个腔内牵张肠发生装置原型的连接部分,以允许模拟肠内容物(SEC)在猪和人空肠中流通。

材料和方法

SEC 使用 80%的玉米糖浆制成。10cm 的猪和人肠段在腔内缝合有一个辐条状的 2.2cm 连接适配器,其流通面积为 33.6mm²。SEC 以 0.083mL/s 的速度泵入肠段的近端。记录 SEC 通过连接器(第一滴)、10mL 和 20mL SEC 从远端洗脱的首次通过时间。

结果

猪肠的首次洗脱第一滴的平均时间为 155 ± 38s,而人肠为 149 ± 22s(P=0.8)。这相当于初始滴通过前的初始静水压为 37.5mmHg。猪肠的平均流速为 0.094mL/s,人肠为 0.084mL/s(P=0.09)。为了考虑到腔内碎片引起的阻塞,在较小的猪肠中研究了连接器孔 75%阻塞的情况,以研究流通面积的减少是否可以耐受。首次洗脱第一滴的平均时间略有增加至 171 ± 15s,但洗脱率保持不变(P=0.5)。

结论

在初始压力升高到允许 SEC 的第一滴通过连接适配器的生理水平后,我们的原型肠连接适配器在流通面积减少 75%的情况下,并没有阻塞 SEC 的流通。这种连接方式代表了一种可行的方法,可以将设备连续放置而不阻塞肠内容物的流动。

相似文献

1
Flow through a mechanical distraction enterogenesis device: a pilot test.经机械牵张成骨器的血流:初步试验。
J Surg Res. 2011 Oct;170(2):179-84. doi: 10.1016/j.jss.2011.03.058. Epub 2011 Apr 19.
2
Spring-mediated distraction enterogenesis in-continuity.弹簧介导的连续性肠再生术。
J Pediatr Surg. 2016 Dec;51(12):1983-1987. doi: 10.1016/j.jpedsurg.2016.09.024. Epub 2016 Sep 15.
5
Scalability of an endoluminal spring for distraction enterogenesis.用于牵引肠再生的腔内弹簧的可扩展性。
J Pediatr Surg. 2016 Dec;51(12):1988-1992. doi: 10.1016/j.jpedsurg.2016.09.023. Epub 2016 Sep 15.
6
Function of mechanically lengthened jejunum after restoration into continuity.恢复连续性后机械延长空肠的功能。
J Pediatr Surg. 2014 Jun;49(6):971-4; discussion 974-5. doi: 10.1016/j.jpedsurg.2014.01.033. Epub 2014 Jan 31.
7
Long-term safety of intraluminal spring-mediated bowel lengthening.腔内弹簧介导的肠道延长术的长期安全性
J Pediatr Surg. 2023 Jan;58(1):89-93. doi: 10.1016/j.jpedsurg.2022.09.034. Epub 2022 Sep 26.
8
Intestinal lengthening in an innovative rodent surgical model.一种创新型啮齿动物手术模型中的肠道延长术
J Pediatr Surg. 2014 Dec;49(12):1791-4. doi: 10.1016/j.jpedsurg.2014.09.022. Epub 2014 Oct 1.

本文引用的文献

1
Application of distractive forces to the small intestine: defining safe limits.牵拉小肠的应用:确定安全限度。
J Surg Res. 2010 Oct;163(2):169-75. doi: 10.1016/j.jss.2010.03.060. Epub 2010 Apr 21.
4
Sustainability of mechanically lengthened bowel in rats.大鼠机械延长肠段的可持续性
J Pediatr Surg. 2006 Dec;41(12):2019-22. doi: 10.1016/j.jpedsurg.2006.08.014.
7
Pediatric short bowel syndrome: redefining predictors of success.小儿短肠综合征:重新定义成功的预测因素。
Ann Surg. 2005 Sep;242(3):403-9; discussion 409-12. doi: 10.1097/01.sla.0000179647.24046.03.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验