Section of Pediatric Surgery, Department of Surgery and the C.S. Mott Children's Hospital, University of Michigan, Ann Arbor, Michigan, USA.
J Surg Res. 2011 Oct;170(2):179-84. doi: 10.1016/j.jss.2011.03.058. Epub 2011 Apr 19.
We tested the coupling portion of a prototype intraluminal distraction enterogenesis device to allow flow-through of simulated enteric contents (SEC) in both pig and human jejunum.
SEC was made using 80% corn syrup. Ten-cm pig and human intestinal segments had a spoke-shaped 2.2 cm coupling adaptor sutured in place, intraluminally. The adaptor had a flow-through area of 33.6 mm(2). SEC was pumped into the proximal part of the intestinal segment at 0.083 mL/s. The times to first passage of SEC through the coupler (first drop), 10 mL, and 20 mL of SEC eluted from the distal end were recorded.
Mean time to first drop elution was 155 ± 38 s with pig, and 149 ± 22 s with human bowel (P = 0.8). This corresponded to a hydrostatic pressure of 37.5 mmHg before the initial drop passed through. Mean flow rates were 0.094 mL/s in pig bowel and 0.084 mL/s in human bowel (P = 0.09). To account for occlusion from luminal debris, a 75% occlusion of coupler holes was studied in the smaller pig bowel to investigate if reductions in flow-through area could be tolerated. Mean time to first drop increased slightly to 171 ± 15 s, but the elution rate stayed the same (P = 0.5).
After a physiologic level of initial pressure buildup allowing the first drop of SEC to pass the coupling adaptor, our prototype intestinal coupling adaptor did not obstruct flow-through of SEC, even after a 75% decrease in flow-through area. This type of attachment represents a viable approach to placing a device in-continuity without obstructing flow of enteric contents.
我们测试了一个腔内牵张肠发生装置原型的连接部分,以允许模拟肠内容物(SEC)在猪和人空肠中流通。
SEC 使用 80%的玉米糖浆制成。10cm 的猪和人肠段在腔内缝合有一个辐条状的 2.2cm 连接适配器,其流通面积为 33.6mm²。SEC 以 0.083mL/s 的速度泵入肠段的近端。记录 SEC 通过连接器(第一滴)、10mL 和 20mL SEC 从远端洗脱的首次通过时间。
猪肠的首次洗脱第一滴的平均时间为 155 ± 38s,而人肠为 149 ± 22s(P=0.8)。这相当于初始滴通过前的初始静水压为 37.5mmHg。猪肠的平均流速为 0.094mL/s,人肠为 0.084mL/s(P=0.09)。为了考虑到腔内碎片引起的阻塞,在较小的猪肠中研究了连接器孔 75%阻塞的情况,以研究流通面积的减少是否可以耐受。首次洗脱第一滴的平均时间略有增加至 171 ± 15s,但洗脱率保持不变(P=0.5)。
在初始压力升高到允许 SEC 的第一滴通过连接适配器的生理水平后,我们的原型肠连接适配器在流通面积减少 75%的情况下,并没有阻塞 SEC 的流通。这种连接方式代表了一种可行的方法,可以将设备连续放置而不阻塞肠内容物的流动。