Rotating disk sorbent extraction for pre-concentration of chromogenic organic compounds and direct determination by solid phase spectrophotometry.

机构信息

Departamento de Química Inorgánica y Analítica, Universidad de Chile, Santiago, Chile.

出版信息

Anal Chim Acta. 2011 Jun 10;695(1-2):73-6. doi: 10.1016/j.aca.2011.04.009. Epub 2011 Apr 16.

Abstract

A novel and very simple microextraction approach for pre-concentration and direct solid phase spectrophotometric measurement has been developed for the determination of chromogenic analytes. The model analyte to assess this approach was the chromophore malachite green (MG). The analyte was extracted from water samples onto a small rotating disk made of Teflon containing a sorbent phase of polydimethylsiloxane (PDMS) on one of its surfaces. We refer to the extraction procedure as rotating disk sorptive extraction (RDSE). After extraction, the sorbent phase with the concentrated analyte was separated from the Teflon disk and used directly for MG determination by solid phase spectrophotometry at 624 nm, without the necessity of a desorption step. Chemical and extraction variables such as concentration of sodium sulfate, pH, disk rotational velocity, extraction time, and temperature were studied in order to establish the best conditions for extraction. Under optimum conditions, the extraction of MG was carried out in 18 min and 90 min, for sample volumes of 100mL or 1000 mL, respectively. The detection limit, based on three times the standard deviation of the blank phase (3σ(b)), was 1.4 μg L⁻¹ and the repeatability, expressed as relative standard deviation (RSD), for 20 μg L⁻¹ MG was 8.1%. This study also applied the method to real samples, obtaining quantitative recovery (mean recovery of 99.3%). The PDMS phases could be reused after desorbing the MG into methanol for 3h. Replacement of the PDMS film onto the disk is very easy and low cost.

摘要

一种新颖且非常简单的微萃取方法已被开发用于预浓缩和直接固相分光光度法测量,以测定显色分析物。评估该方法的模型分析物是显色剂孔雀石绿 (MG)。将分析物从水样中萃取到一个小的旋转圆盘上,该圆盘由聚四氟乙烯制成,其一个表面上含有聚二甲基硅氧烷 (PDMS) 的吸附剂相。我们将萃取程序称为旋转圆盘吸附萃取 (RDSE)。萃取后,将含有浓缩分析物的吸附剂相与聚四氟乙烯圆盘分离,并直接用于固相分光光度法在 624nm 处测定 MG,无需解吸步骤。研究了化学和萃取变量,例如硫酸钠浓度、pH 值、圆盘旋转速度、萃取时间和温度,以确定最佳萃取条件。在最佳条件下,对于 100mL 或 1000mL 样品体积,MG 的萃取分别在 18 分钟和 90 分钟内完成。基于空白相三倍标准偏差 (3σ(b)) 的检测限为 1.4μg L⁻¹,对于 20μg L⁻¹ MG 的重复性表示为相对标准偏差 (RSD),为 8.1%。该研究还将该方法应用于实际样品,获得了定量回收(MG 的平均回收率为 99.3%)。在将 MG 解吸到甲醇中 3 小时后,可以重复使用 PDMS 相。将 PDMS 薄膜更换到圆盘上非常容易且成本低。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验