Walling Leslie A
Amgen, Inc., M/S 29-1-A, One Amgen Center Drive, Thousand Oaks, CA 91320, USA.
J Lab Autom. 2011 Jun;16(3):235-40. doi: 10.1016/j.jala.2011.01.001. Epub 2011 May 16.
Serial dilution of compounds solubilized in dimethylsulfoxide (DMSO) for dose-response curves is a common method for efficacy analysis of potential drug candidates. In general, serial dilution methods are particularly prone to error propagation because each dilution is dependent on the previous concentration. Moreover, assumptions about quality control parameters (i.e., dye linearity) can lead to an erroneous process. Here, an inline performance measurement is sought to improve the precision and accuracy of dilution plates. Sulforhodamine 101 (S101) dye is introduced as the quantitative fluorometric method of choice for DMSO-based systems. Although S101 in DMSO behaves in a nonlinear fashion over its detectable range, we account for this with a direct calibration method that includes every point of the dilution template. This report contains dye selection rationale for the S101 dye and its use in quantifying the performance of 96- and 384-well dilution protocols as tested on five identical instruments.