Shin J H, Sakong D S, Nam H, Cha G S
Department of Chemistry and Research Institute of Basic Science, Kwangwoon University, 447-1 Wolgye-Dong, Nowon-Ku, Seoul 139-701, Korea.
Anal Chem. 1996 Jan 1;68(1):221-5. doi: 10.1021/ac9506037.
A new silicone rubber matrix carbonate-selective membrane and a high-pH buffer diluent are used to enhance the performance of the electrode measurements for serum carbon dioxide. The proposed membrane employs one-component silicone rubber as the matrix and trifluoroacetyl-p-decylbenzene as the neutral ionophore. The optimized membrane formulation incorporates as high as 21.9 wt % plasticizer (e.g., bis(2-ethylhexyl) adipate). The highly plasticized silicone rubber membranes not only function equivalently, in terms of the carbonate response, to the conventional PVC matrix membranes, but they also exhibit substantially reduced interfering response toward salicylate. Furthermore, the silicone rubber membrane exhibits better adhesion to the solid surface than do PVC or PU membranes. The use of higher pH buffers (e.g., 2-amino-2-methyl-1-propanol (AMP)-H(2)SO(4), pH 9.5-10.5) further enhances the selectivity of the carbonate electrode measurement system for total CO(2) species over other anions. It is shown that the combined use of the silicone rubber matrix membrane and the high-pH AMP buffer provides a carbonate sensor system that is substantially less subject to interference from salicylate and chloride than is the conventional measurement system employing the PVC-based electrode with the lower pH (8.4-8.8) buffer diluent.
一种新型的硅橡胶基质碳酸根选择性膜和一种高pH缓冲稀释剂被用于提高血清二氧化碳电极测量的性能。所提出的膜采用单组分硅橡胶作为基质,三氟乙酰对癸基苯作为中性离子载体。优化后的膜配方含有高达21.9 wt%的增塑剂(如己二酸二(2-乙基己基)酯)。高度增塑的硅橡胶膜在碳酸根响应方面不仅与传统的PVC基质膜功能相当,而且对水杨酸酯的干扰响应也大幅降低。此外,硅橡胶膜对固体表面的附着力比PVC或PU膜更好。使用更高pH的缓冲剂(如2-氨基-2-甲基-1-丙醇(AMP)-H₂SO₄,pH 9.5 - 10.5)进一步提高了碳酸根电极测量系统对总CO₂物种相对于其他阴离子的选择性。结果表明,与采用低pH(8.4 - 8.8)缓冲稀释剂的PVC基电极的传统测量系统相比,硅橡胶基质膜和高pH的AMP缓冲剂的联合使用提供了一种基本上较少受到水杨酸酯和氯离子干扰的碳酸根传感器系统。