Servicio de Radiofísica, Hospital Universitario San Cecilio, Avda Dr Olóriz, 16, E-18012 Granada, Spain.
Phys Med. 2012 Jul;28(3):254-61. doi: 10.1016/j.ejmp.2011.05.002. Epub 2011 Jun 8.
We aimed to analyze the optimal conditions to carry out the periodical statistical control tests of the gamma camera count rate performance. First we focused in reproducing the actual R(20) value of the gamma camera response. Second we studied the variability of this parameter in the statistical control test. We performed a reference measurement, which consisted of the determination of the complete curve relating observed and incident count rates, the counting model describing it and the reference R(20). This reference determined the conditions for the statistical control tests and the way to analyze the results obtained. Results from three different gamma cameras were studied. Each gamma camera showed a different behavior and required specific data analysis. The optimal conditions to perform the statistical control test were determined in each case. Our procedure provides the information necessary to correlate the average value of R(20) obtained in the quality control test with the reference one. The critical requirement to perform any statistical control test, that is to have a reduced variability of the control variable, can be fulfilled in this case only for relatively high activities.
我们旨在分析伽玛相机计数率性能定期统计控制测试的最佳条件。首先,我们专注于再现伽玛相机响应的实际 R(20)值。其次,我们研究了该参数在统计控制测试中的可变性。我们进行了参考测量,包括确定观察到的和入射计数率之间的完整曲线、描述它的计数模型和参考 R(20)。该参考确定了统计控制测试的条件和分析获得结果的方法。研究了来自三个不同伽玛相机的数据。每个伽玛相机都表现出不同的行为,需要特定的数据分析。在每种情况下都确定了执行统计控制测试的最佳条件。我们的程序提供了将质量控制测试中获得的 R(20)平均值与参考值相关联所需的信息。执行任何统计控制测试的关键要求,即控制变量的可变性降低,在这种情况下仅在相对较高的活度下才能满足。