Division of Laboratory Science, National Center for Environmental Health, Centers for Disease Control and Prevention, Atlanta, GA 30341, USA.
J Chromatogr B Analyt Technol Biomed Life Sci. 2011 Jul 15;879(22):2142-8. doi: 10.1016/j.jchromb.2011.05.054. Epub 2011 Jun 12.
Nicotine (NIC), cotinine (COT) and trans-3'-hydroxycotinine (OHCOT) are the most prevalent and abundant tobacco biomarkers in meconium. We have developed and validated an accurate and precise method for the measurement of these analytes in meconium in which potassium hydroxide is used to digest the meconium sample, followed by solid phase extraction from the liquified sample. The precision of OHCOT, COT and NIC measurements (intra-day and inter-day) were 4.8-10.6%, 3.4-11.6% and 9.3-15.8%, respectively. Evaluation of accuracy indicated bias of -4.0, 2.0 and 0.8% for OHCOT at concentrations of 0.5, 2.5 and 7.5 ng/g. The accuracy estimates for COT at concentrations of 0.5, 2.5 and 7.5 ng/g are 4.0, 4.0 and 5.7%, respectively. For NIC at 2, 10 and 30 ng/g the accuracy was calculated to be 3.0, 5.0 and 5.1%, respectively. The linear range of standard solutions was 0.125-37.5 ng/mL for OHCOT and COT, and 0.75-150 ng/mL for NIC. This method was applied to the analysis of 374 meconium samples from infants of both smoking and nonsmoking mothers. Positive correlations with r(2)≥0.63 were observed between NIC and COT, COT and OHCOT, NIC and OHCOT, and NIC and (OHCOT+COT) in these samples.
尼古丁(NIC)、可替宁(COT)和反式-3'-羟基可替宁(OHCOT)是胎粪中最普遍和丰富的烟草生物标志物。我们已经开发并验证了一种准确而精密的方法,用于测量胎粪中的这些分析物,其中使用氢氧化钾消化胎粪样品,然后从液化样品中进行固相萃取。OHCOT、COT 和 NIC 测量(日内和日间)的精密度分别为 4.8-10.6%、3.4-11.6%和 9.3-15.8%。准确度评估表明,在 0.5、2.5 和 7.5 ng/g 浓度下,OHCOT 的偏倚为-4.0、2.0 和 0.8%。在 0.5、2.5 和 7.5 ng/g 浓度下,COT 的准确度估计值分别为 4.0、4.0 和 5.7%。在 2、10 和 30 ng/g 浓度下,NIC 的准确度计算值分别为 3.0、5.0 和 5.1%。标准溶液的线性范围为 0.125-37.5 ng/mL 用于 OHCOT 和 COT,0.75-150 ng/mL 用于 NIC。该方法应用于分析 374 份来自吸烟和不吸烟母亲的婴儿胎粪样本。在这些样本中,观察到 NIC 与 COT、COT 与 OHCOT、NIC 与 OHCOT 以及 NIC 与(OHCOT+COT)之间的相关性 r(2)≥0.63。