Service de diabétologie, hôpital Haut-Lévêque, 33604 Pessac, France.
Diabetes Metab. 2011 Dec;37(6):527-32. doi: 10.1016/j.diabet.2011.05.003. Epub 2011 Jun 28.
Perturbation of pain sensation is considered one of the major initiating risk factors for diabetic foot ulcer. Sweat dysfunction leading to abnormal skin conditions, including dryness and fissures, can increase foot ulcer risk. The aim of this study was to evaluate Sudoscan™, a new, quick, non-invasive and quantitative method of measuring sudomotor dysfunction as a co-indicator of the severity of diabetic polyneuropathy (DPN).
A total of 142 diabetic patients (age 62±18 years, diabetes duration 13±14 years, HbA(1c) 8.9±2.5%) were measured for vibration perception threshold (VPT), using a biothesiometer, and for sudomotor dysfunction, using electrochemical sweat conductance (ESC) based on the electrochemical reaction between sweat chloride and electrodes in contact with the hands and feet. Retinopathy status was also assessed, as well as reproducibility between two ESC measurements and the effect of glycaemia levels.
ESC measurements in the feet of patients showed a descending trend from 66±17 μS to 43±39 μS, corresponding to an ascending trend in VPT threshold from <15 V to >25 V (P=0.001). Correlation between VPT and ESC was -0.45 (P<0.0001). Foot ESC was lower in patients with fissures, while VPT was comparable. Both VPT and foot ESC correlated with retinopathy status. Bland-Altman plots indicated good reproducibility between two measurements, and between low and high glycaemia levels.
Sudoscan™ is a reproducible technique with results that are not influenced by blood glucose levels. Sweating status may be a quantitative indicator of the severity of polyneuropathy that may be useful for the early prevention of foot skin lesions.
痛觉障碍被认为是糖尿病足溃疡的主要起始风险因素之一。出汗功能障碍导致皮肤异常,包括干燥和皲裂,会增加足溃疡的风险。本研究旨在评估 Sudoscan™,这是一种新的、快速、非侵入性和定量的测量汗分泌功能障碍的方法,作为糖尿病多发性神经病(DPN)严重程度的共同指标。
共测量了 142 名糖尿病患者(年龄 62±18 岁,糖尿病病程 13±14 年,HbA1c 8.9±2.5%)的振动感觉阈值(VPT),使用生物感觉计,以及使用电化学汗导率(ESC)测量汗分泌功能障碍,基于电化学汗液中的氯与电极之间的电化学反应,电极与手和脚接触。还评估了视网膜病变状态,以及两次 ESC 测量之间的重复性以及血糖水平的影响。
患者足部的 ESC 测量值从 66±17 μS 下降到 43±39 μS,相应的 VPT 阈值从<15 V 上升到>25 V(P=0.001)。VPT 和 ESC 之间存在负相关(r=-0.45,P<0.0001)。有皲裂的患者足部 ESC 较低,而 VPT 则相似。VPT 和足部 ESC 均与视网膜病变状态相关。Bland-Altman 图表明两次测量之间以及高低血糖水平之间具有良好的可重复性。
Sudoscan™ 是一种可重复的技术,其结果不受血糖水平的影响。出汗状态可能是多发性神经病严重程度的定量指标,有助于早期预防足部皮肤病变。