Sinars D B, Wenger D F, Pikuz S A, Jones B, Geissel M, Hansen S B, Coverdale C A, Ampleford D J, Cuneo M E, McPherson L A, Rochau G A
Sandia National Laboratories, P.O. Box 5800, Albuquerque, New Mexico 87185, USA.
Rev Sci Instrum. 2011 Jun;82(6):063113. doi: 10.1063/1.3600610.
We describe a pair of time-integrated transmission spectrometers that are designed to survey 7-28 keV (1.9 to 0.43 Å) x-ray photons produced by experiments on the Sandia Z pulsed power facility. Each spectrometer uses a quartz 10-11 crystal in a Cauchois geometry with a slit to provide spatial resolution along one dimension. The spectrometers are located in the harsh environment of the Z vacuum chamber, which necessitates that their design be compact and rugged. Example data from calibration tests and Z experiments are shown that illustrate the utility of the instruments.