Affatato Saverio, Traina Francesco, Toni Aldo
Medical Technology Laboratory, Rizzoli Orthopedic Institute, Bologna, Italy.
Int J Artif Organs. 2011 Jun;34(6):506-12. doi: 10.5301/IJAO.2011.8457.
The combination of materials that still has highest wear resistance for total hip replacement is ceramic-on-ceramic. However, brittleness is a major concern for ceramics: in vivo and in vitro studies on ceramic hip prostheses correlate microseparation with hip noise, ceramic wear, or ceramic liner damage. Ceramic microseparation can lead to edge load, ceramic head wear, and squeaking. The aim of this in vitro study was to investigate whether different angles of inclination influence the wear pattern of alumina-on-alumina hip joints with micro-separation during the swing phase. We also evaluated the wear rate obtained from this in vitro investigation with retrieval specimens obtained at 13 years' mean follow-up. The study was performed using a 12-station hip joint wear simulator (Shore Western, Monrovia, CA, USA) under bovine calf serum used as lubricant. Wear was evaluated by the gravimetric method and the test length was set at two million cycles. After two million cycles, a volumetric loss of 0.11 ±0.03 mm3 and 0.12 ±0.06 mm3 was observed, respectively, for 23° and 63° angles of inclination. In particular, the results obtained in this work revealed an increase of about 12-fold compared to previous results without microseparation conditions. No significant differences were observed between the two different inclinations on the wear patterns of the acetabular cups with a level of significance of a = 0.5. The location and general shape of the stripes wear were similar for the retrieved and simulator balls.
对于全髋关节置换而言,具有最高耐磨性的材料组合是陶瓷对陶瓷。然而,脆性是陶瓷的一个主要问题:对陶瓷髋关节假体的体内和体外研究表明,微分离与髋关节异响、陶瓷磨损或陶瓷内衬损坏有关。陶瓷微分离会导致边缘负荷、陶瓷股骨头磨损和吱吱声。这项体外研究的目的是调查不同倾斜角度是否会影响在摆动阶段存在微分离的氧化铝对氧化铝髋关节的磨损模式。我们还将此次体外研究获得的磨损率与在平均13年随访时获取的翻修标本的磨损率进行了评估。该研究使用一台12工位的髋关节磨损模拟器(美国加利福尼亚州蒙罗维亚的肖尔西部公司),以小牛血清作为润滑剂。通过重量法评估磨损情况,试验长度设定为两百万次循环。两百万次循环后,对于23°和63°的倾斜角度,分别观察到体积损失为0.11±0.03立方毫米和0.12±0.06立方毫米。特别是,与之前无微分离条件下的结果相比,这项工作所获得的结果显示增加了约12倍。在磨损模式方面,对于髋臼杯,在显著性水平α = 0.5时,两种不同倾斜角度之间未观察到显著差异。翻修球和模拟器球的条纹磨损位置和总体形状相似。