J Med Eng Technol. 2011 Aug-Oct;35(6-7):330-7. doi: 10.3109/03091902.2011.592237. Epub 2011 Jul 8.
A core defect in people with Type 2 Diabetes is endothelial dysfunction. This defect permeates all organ systems in the body including the ability of the skin to protect itself from thermal injuries by an appropriate increase in skin circulation. Most studies on the local response to heating have been done with dry heat sources. Recent data show that endothelial function is improved in people with diabetes with moist heat. Little is known about 'how' moist heat must be or the mechanisms on why moist heat triggers a better blood flow response than dry heat.
In the present investigation, a device was developed to provide variable temperature air and variable humidity as an aid to study the dynamics of the skin circulatory response to heat in people with diabetes. The device consisted of a water bath used to heat air and an air dryer and air bubbler to generate dry and moist air, respectively, at a fixed temperature. The air could then be mixed and the temperature stabilized to produce a variable temperature and humidity air source to expose the skin to in people with diabetes.
The device was validated at different air temperatures and humidities and tested on four subjects to assess operation. The air flows, temperatures and humilities were stable with less than a 5% coefficient of variation.
Testing on humans showed that there appeared to be a linear relationship between air humidity and blood flow at a given air temperature exposed to the skin.
2型糖尿病患者的一个核心缺陷是内皮功能障碍。这种缺陷影响身体的所有器官系统,包括皮肤通过适当增加皮肤循环来保护自身免受热损伤的能力。大多数关于局部受热反应的研究都是使用干热源进行的。最近的数据表明,湿热可改善糖尿病患者的内皮功能。对于湿热必须达到何种程度,以及湿热引发比干热更好的血流反应的机制,人们知之甚少。
在本研究中,开发了一种装置,用于提供可变温度的空气和可变湿度,以帮助研究糖尿病患者皮肤对热的循环反应动态。该装置由一个用于加热空气的水浴、一个空气干燥器和一个空气鼓泡器组成,分别用于在固定温度下产生干燥空气和潮湿空气。然后可以将空气混合并稳定温度,以产生可变温度和湿度的空气源,用于让糖尿病患者的皮肤暴露于其中。
该装置在不同的空气温度和湿度下进行了验证,并在四名受试者身上进行了测试以评估其运行情况。气流、温度和湿度稳定,变异系数小于5%。
人体测试表明,在给定的暴露于皮肤的空气温度下,空气湿度与血流之间似乎存在线性关系。