Chemistry Research Laboratory, Department of Chemistry, University of Oxford, Oxford OX1 3TA, United Kingdom.
J Chromatogr A. 2011 Aug 12;1218(32):5480-6. doi: 10.1016/j.chroma.2011.06.044. Epub 2011 Jun 17.
Hydrophilic interaction liquid chromatography (HILIC) is here successfully coupled to negative-ion electrospray ionization time-of-flight mass spectrometry (ESI-TOFMS) for the analysis of synthetic and chemically modified oligonucleotides. Separation was performed on a 2.1 mm × 100 mm PEEK ZIC HILIC column packed with hydrophilic stationary phase with a permanent zwitterionic functional group and a particle size of 3.5 μm with an average pore diameter of 200Å. A method was developed to separate homogeneous and heterogeneous oligonucleotides as well as methylated oligonucleotides using a quaternary pumping system containing ammonium acetate and water with an acetonitrile gradient. Analyses of oligonucleotides were performed by LC/MS with a detection limit of 2.5 picomole (20 mer) with signal to noise ratio (S/N) of 4.12. The influence of the eluent composition, type of buffer and its concentration, and organic modifier were also evaluated. The HILIC LC/MS method presented in this paper used common, 'MS friendly', mobile phases achieving sensitive and selective oligonucleotide analysis.
亲水作用色谱(HILIC)与负离子电喷雾飞行时间质谱(ESI-TOFMS)成功联用,用于分析合成和化学修饰的寡核苷酸。分离在填充有亲水固定相的 2.1mm×100mm PEEK ZIC HILIC 柱上进行,亲水固定相带有永久两性离子官能团,粒径为 3.5μm,平均孔径为 200Å。开发了一种使用含有乙酸铵和水的四元泵送系统以及乙腈梯度来分离均相和异相寡核苷酸以及甲基化寡核苷酸的方法。通过 LC/MS 分析,寡核苷酸的检测限为 2.5 皮摩尔(20 个碱基对),信噪比(S/N)为 4.12。还评估了洗脱液组成、缓冲液类型及其浓度以及有机溶剂的影响。本文提出的 HILIC LC/MS 方法使用了常见的、“MS 友好”的流动相,实现了对寡核苷酸的灵敏和选择性分析。